Performance characteristics of creatine kinase-MB isoenzyme measured with an immunoenzymometric and an immunoinhibition assay in acute myocardial infarction with and without thrombolytic therapy.
نویسندگان
چکیده
In a time study we compared the analytical and clinical performance of the Tandem Icon QSR CK-MB enzyme-immunoassay (Hybritech) (creatine kinase-MB) and a creatine kinase-MB immunoinhibition method (Boehringer Mannheim GmbH) (creatine kinase-B). Two hundred and ninety-nine serum samples from 38 patients suspected of acute myocardial infarction were collected at regular intervals during 48 hours. Twenty-nine patients were diagnosed as having acute myocardial infarction, of whom 19 received thrombolytic therapy. Although highly correlated, the large scatter around the regression line at low values indicated a different clinical performance of the two methods. We evaluated and compared test performance at different decision levels by means of frequency distributions and predictive values of positive and negative results. For early diagnosis of acute myocardial infarction (4 hours after onset of pain) the Hybritech creatine kinase-MB method gives acceptable predictive values. In thrombolytic treated acute myocardial infarction patients, the peak creatine kinase-MB and creatine kinase-B concentrations were reached after 13.0 h and 13.6 h after the onset of pain, compared with 19.8 h and 17.8 h for patients without thrombolytic therapy.
منابع مشابه
Mass concentration and activity concentration of creatine kinase isoenzyme MB compared in serum after acute myocardial infarction.
We compared three current methods (immunoinhibition, "Isomune-CK" immunoprecipitation, and the Tandem-E CKMB II immunoenzymometric assay) for determination of creatine kinase (CK; EC 2.7.3.2) isoenzyme MB in serum. Although results inter-correlated well, the immunoinhibition assay gave higher activity values. Atypical CK forms did not interfere with the immunoprecipitation and immunoenzymometri...
متن کاملAnalytical and clinical evaluation of creatine kinase MB mass assay by IMx: comparison with MB isoenzyme activity and serum myoglobin for early diagnosis of myocardial infarction.
We analytically and clinically evaluated Abbott's IMx assay for creatine kinase (CK) isoenzyme MB (CK-MB) in serum. Over a 1-year period, the method was more specific but less precise than catalytic isoenzyme measurements by electrophoresis or immunoinhibition. Sera from different individuals without electrophoretic evidence of CK-MB but containing macro CK type 1 (n = 20), mitochondrial CK (n ...
متن کاملEffect of the Inhibition of Mitochondrial Creatine Kinase Activity on the Clinical Diagnosis of Suspected Acute Myocardial Infarction
Background: Creatine kinase isoenzyme MB (CK-MB) activity using the immunoinhibition method has been measured in clinical laboratories. However, the presence of mitochondrial CK (MtCK) abnormally increases CK-MB activity when measured using the conventional method, which can lead to misdiagnosis of acute myocardial infarction (AMI). A new CK-MB activity assay kit that can inhibit MtCK activity ...
متن کاملEarly diagnosis of acute myocardial infarction by a newly developed rapid immunoturbidimetric assay for myoglobin.
OBJECTIVE To evaluate a rapid immunoturbidimetric assay for myoglobin and to investigate its clinical usefulness in the early detection of acute myocardial infarction. DESIGN Prospective study. Immunoturbidimetrically determined myoglobin concentrations were compared with radioimmunoassay results obtained with the same blood samples. The diagnostic performance of myoglobin determination was c...
متن کاملActivity concentration and mass concentration (monoclonal antibody immunoenzymometric method) compared for creatine kinase MB isoenzyme in serum.
Results of the "Tandem-E CKMB" immunoenzymometric procedure (y) for creatine kinase (CK; EC 2.9.3.2) were compared with electrophoresis (x) for 160 serum samples from patients suspected of having sustained myocardial infarctions. The results correlated well: y, microgram/L (Tandem assay) = 1.3x-6.3 U/L(electrophoresis) (r = 0.95). CK-MB mass measurement was more stable than enzyme activity afte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- European journal of clinical chemistry and clinical biochemistry : journal of the Forum of European Clinical Chemistry Societies
دوره 30 6 شماره
صفحات -
تاریخ انتشار 1992