Power vector analysis as an aid to correct a rotated Artiflex toric phakic intraocular lens
نویسنده
چکیده
We describe the use of power vector analysis in quantifying the amount of lens rotation and assesing the subsequent correction in a case of rotated Artiflex toric phakic intraocular lens (pIOL) (Ophtec BV, Netherlands). A 33 year old patient was corrected by bilateral implantation of Artiflex toric pIOLs for high myopia and astigmatism. Two months after the surgery the patient reported worse vision with her left eye, with residual refractive error of +1.25 –2.00 x 135° yielding a corrected visual acuity of 1.0. Pre-operative and post-operative refractions in vectorial notation were subtracted and the result changed back into clinical notation, the resulting refraction being –8.98 –4.03 x 15°, which was the same amount as attempted but rotated 15 degrees counterclockwise. A second procedure to rotate the position of the lens clockwise by this amount resulted in a residual error of +0.50 –0.75 x 105° and uncorrected visual acuity of 1.0–.
منابع مشابه
Vector analysis of corneal and refractive astigmatism changes following toric pseudophakic and toric phakic IOL implantation.
PURPOSE To determine the efficacy of the astigmatism correction following toric intraocular lens (IOL) and toric phakic IOL (pIOL) implantation in eyes with no previous ocular surgery and in postkeratoplasty (PKP) eyes. In addition, changes in corneal astigmatism were determined. METHODS Astigmatism was analyzed in 35 eyes with an AcrySof toric IOL, 35 eyes with an Artiflex toric pIOL, 50 eye...
متن کاملArtiflex Toric Phakic Intraocular Lens Implantation in Congenital Nystagmus
CASE A 44-year-old woman with congenital nystagmus and myopic astigmatism in both eyes who was submitted to phakic intraocular lens (pIOL) implantation. METHODS Full ophthalmologic examination including refractive status, corrected (CDVA) and uncorrected (UCVA) monocular and binocular visual acuities, ocular motility, slit-lamp evaluation, tonometry and fundoscopy before and after implantatio...
متن کاملFlexible Toric Iris Claw Phakic Intraocular Lens Implantation for Myopia and Astigmatism
PURPOSE To assess the visual and refractive outcomes of flexible toric iris claw phakic intraocular lens implantation for correction of moderate to high myopia with astigmatism. METHODS In this non-randomized prospective study, 31 eyes of 18 patients including 8 male and 10 female subjects with mean age of 27.62±5.53 (range 19 to 38) years with moderate to high myopia and astigmatism underwen...
متن کاملTraumatic Partial Aniridia and Cataract after Iris-Fixated Foldable Phakic Intraocular Lens Implantation
We present the first case of partial aniridia, traumatic cataract, and peripapillary subretinal hemorrhage after blunt trauma to an eye containing a foldable iris-fixated phakic intraocular lens (pIOL). Although the patient experienced visual loss because of considerable corneal astigmatism and a traumatic cataract, vision was recovered by a 2-stage primary corneal repair and cataract surgery w...
متن کاملEndothelial distance after phakic iris-fixated intraocular lens implantation: a new safety reference
PURPOSE To evaluate the distance between the endothelial surface of the cornea to the anterior edge of an Artiflex® phakic intraocular lens (IOL) implant to improve the safety profile of this implant. METHODS This is a retrospective clinical case series of 45 patients who had Artiflex phakic IOL implantation (Artiflex p-IOL) with a follow-up period of 3 years. A Pentacam HR imaging system was...
متن کامل