Case Report of an open Neck Procedure Complication Associated with Laryngeal Mask Airway Use
نویسندگان
چکیده
Introduction The laryngeal mask airway (LMA) is a safe method to establish airway control during general anaesthesia. In otolaryngology surgery, the use of a LMA is well established in ear surgery; however, the use of a LMA during open neck procedures remains controversial. We report a case in which the pharyngeal distortion by the LMA cuff resulted in an iatrogenic complication. Case Report A 38-year-old female with a background of multiple myeloma was referred to the otolaryngology team for an open cervical lymph node biopsy. The patient was in remission after a 2 year post chemotherapy treatment, but now presented with a 4-week history of persistent nodal enlargement. During the elective procedure, pharyngeal distortion from the laryngeal airway mask used for airway management resulted in an iatrogenic pharyngeal injury. This case is reported to highlight the importance of communication between the surgeon and anesthetist about the mode of airway management in open neck surgery. Conclusion Communication between the otolaryngologist and anesthetist is pertinent when selecting the method of airway management in open neck procedures. A LMA should be used with caution during open neck procedures, with the surgeon recognizing the potential for pharyngeal distortion.
منابع مشابه
Case Report of an open Neck Procedure Complication Associated with Laryngeal Mask Airway Use
Introduction: The laryngeal mask airway (LMA) is a safe method to establish airway control during general anaesthesia. In otolaryngology surgery, the use of a LMA is well established in ear surgery; however, the use of a LMA during open neck procedures remains controversial. We report a case in which the pharyngeal distortion by the LMA cuff resulted in an iatrogenic complication. Case Report...
متن کاملLimitation of Laryngeal Mask Airway Usage in Open-Neck Procedure
I read the manuscript entitled “Case Report of An Open Neck Procedure Complication Associated with Laryngeal Mask Airway Use” by Subramaniam T et al. This is a valuable manuscript describing iatrogenic complications during level-2 cervical lymph node biopsy, in which the airway is managed with laryngeal mask airway (LMA). During the procedure, the surgeon realized that the mass was actually the...
متن کاملTapia's Syndrome after Corrective Jaw Surgery under General Anesthesia: A Case Report
Introduction:Tapia’s syndrome is a rare complication of recurrent laryngeal and hypoglossal nerve paralysis due to anesthetic airway mismanagement or malpositioning of the patient’s head during surgery. Case Report:Here we present a case of Tapia's syndrome in a 22-year-old male after corrective jaw surgery under general anesthesia, with a long period of recovery, related to airway management p...
متن کاملInferior alveolar nerve injury with laryngeal mask airway: a case report
INTRODUCTION The incidence of damage to the individual cranial nerves and their branches associated with laryngeal mask airway use is low; there have been case reports of damage to the lingual nerve, hypoglossal nerve and recurrent laryngeal nerve. To the best of our knowledge we present the first reported case of inferior alveolar nerve injury associated with laryngeal mask airway use. CASE ...
متن کاملComparison of the Incidence of Sore Throat and Hemodynamic Changes in Airway Management by Macintosh, Glideoscope Laryngoscope and Laryngeal Mask Airway Techniques: Randomized Controlled Trial
Background: Studies have shown that postoperative sore throat is the most common complication of tracheal intubation, However, the actual incidence and extent of it are not well documented. Objective: The aim of this study was to evaluate and compare the incidence of sore throat and hemodynamic Changes in Macintosh, Glidescope Laryngoscope and laryngeal mask airway techniques in elective surg...
متن کامل