Reducing the radiation sterilization dose improves mechanical and biological quality while retaining sterility assurance levels of bone allografts.

نویسندگان

  • Huynh Nguyen
  • Alan I Cassady
  • Michael B Bennett
  • Evelyne Gineyts
  • Andy Wu
  • David A F Morgan
  • Mark R Forwood
چکیده

BACKGROUND Bone allografts carry a risk of infection, so terminal sterilization by gamma irradiation at 25kGy is recommended; but is deleterious to bone quality. Contemporary bone banking significantly reduces initial allograft bioburden, questioning the need to sterilize at 25kGy. METHODS We inoculated allograft bone with Staphylococcus epidermidis and Bacillus pumilus, then exposed them to gamma irradiation at 0, 5, 10, 15, 20 and 25kGy. Mechanical and biological properties of allografts were also assessed. Our aim was to determine an optimal dose that achieves sterility assurance while minimizing deleterious effects on allograft tissue. RESULTS 20-25kGy eliminated both organisms at concentrations from 10(1) to 10(3)CFU, while 10-15kGy sterilized bone samples to a bioburden concentration of 10(2)CFU. Irradiation did not generate pro-inflammatory bone surfaces, as evidenced by macrophage activation, nor did it affect attachment or proliferation of osteoblasts. At doses ≥10kGy, the toughness of cortical bone was reduced (P<0.05), and attachment and fusion of osteoclasts onto irradiated bone declined at 20 and 25kGy (P<0.05). There was no change in collagen cross-links, but a significant dose-response increase in denatured collagen (P<0.05). CONCLUSIONS Our mechanical and cell biological data converge on 15kGy as a threshold for radiation sterilization of bone allografts. Between 5 and 15kGy, bone banks can undertake validation that provides allografts with an acceptable sterility assurance level, improving their strength and biocompatibility significantly. CLINICAL RELEVANCE The application of radiation sterilization doses between 5 and 15kGy will improve bone allograft mechanical performance and promote integration, while retaining sterility assurance levels. Improved quality of allograft bone will promote superior clinical outcomes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Validation of 11 kGy as a radiation sterilization dose for frozen bone allografts.

A radiation sterilization dose (RSD) of 25 kGy is deleterious to bone allografts. This study aimed to establish a lower RSD for bone allografts using method 1 of International Standard Organisation 11137.2:2006. This provides a database to select an RSD corresponding to an allograft's bioburden, given that the bioburden's gamma resistance is equal to or less than the standard. This can be verif...

متن کامل

Bioburden and theoretical sterility dose calculation for radiation sterilization of surgical cotton and bandages.

The initial bioburden, count of colony forming unit (CFU) was determined on the locally manufactured non-sterilized surgical cotton and bandages. In all 489 tests were conducted on 163 cotton samples and 246 tests on 82 bandage samples. The surgical cotton showed an average of 198 microbes with a maxima of 287 and minima 94 whereas bandages showed an average of 179 microbes with a maxima of 268...

متن کامل

Low-Dose Irradiation of Allografts for Ligament and Tendon Reconstruction

Soft tissue allografts for ligament and tendon reconstruction have been used successfully for many years, and their usage has increased dramatically in the past decade. The rate of soft tissue allograft usage in primary anterior cruciate ligament (ACL) reconstruction vs. autograft usage went up from 17% of all surgeries in 2002 to 46% of all surgeries in 20081. Allografts are either processed a...

متن کامل

Evaluation of Accuracy and Quality assurance of external beam therapy with photons

Introduction: Receiving exact dose by the patients is vital in radiotherapy. In radiation therapy, the dosimetry of radiations is too important because of successful radiation inquires for delivering the exact dose to the target volume. This study is to evaluate the tolerances and the accuracy of calculated dose of photon beams in the treatment software system. The TECDOC1583 p...

متن کامل

Mechanical strength of cortical allografts with gamma radiation, in Acta Orthop Belg

versus ethylene oxide (ETO) sterilization on the mechanical strength of cortical bone grafts. Tibias were collected from cadavers of mature goats. Sixty test specimens were randomized into four groups : fresh (no processing), frozen (freezing at –70°C), gamma-irradiated, and ETO-sterilized specimens. Torsion, three-point bending, and compression testing were separately performed with a material...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Bone

دوره 57 1  شماره 

صفحات  -

تاریخ انتشار 2013