A Planar Interdigital Sensor for Bio-impedance Measurement: Theoretical analysis, Optimization and Simulation
نویسندگان
چکیده
This paper proposes the design of a biosensor to characterize the dielectric and conductive properties of biological materials (for example blood or water) by impedance spectroscopy. Particularly, its design optimized the geometric structure interdigitated electrodes. This optimization allows extending the frequency range of measurement by reducing the polarization effect. Polarization effect is manifested by an interface capability (or double layer) from interaction between ions and molecules in the boundary between the surface of the electrolyte and the electrodes, it increases the measurement error at low frequencies. This paper recommends also a novel method to determine the parameters (relative permittivity, thickness and capacitance per unit area) of the double layer (DL) at the contact surface of the electrode with the solution. CoventorWare software was utilized to modelize of interdigital sensor structure in three dimensions (3D) to verify the analytical results and evaluate the influence of geometrical parameters and the dielectric properties of the medium on bioelectrical impedance.
منابع مشابه
Geometric parameters optimization of planar interdigitated electrodes for bioimpedance spectroscopy
This paper is concerned with a physical model of an interdigitated sensor working in a frequency range from 100 Hz to 10 MHz. A theoretical approach is proposed to optimize the use of the sensor for bioimpedance spectroscopy. The correlation between design parameters and frequency behavior in coplanar impedance sensors are described. CoventorWare software was used to model the biological medium...
متن کاملPortable and Precise Measurements with Interdigital Electrodes at Wide Range of Conductivity
A planar structure of the electrodes employed in conductivity measurements is utilizable at small range of conductivity measurement because of small sizes, planar ordering and technology which significantly influence the impedance behavior. The paper presents new approach in the field of precise electrolytic conductivity measurements with planar sensors. This novel measuring method was suggeste...
متن کاملPhase Angle Measurement in Healthy Human Subjects through Bio-Impedance Analysis
Objective(s)Bioelectrical impedance is the measure of impedance of the body. Impedance consists of electric resistance and reactance. Phase angle (PA) is the tan value of the ratio of reactance versus electric resistance. PA depends on cell membrane integrity and on body cell mass. There exists a correlation between PA values and body cell mass.The objective of this study was to compare the PA ...
متن کاملDesign of interdigital spiral and concentric capacitive sensors for materials evaluation
This paper describes the design of two circular coplanar interdigital sensors with i) a spiral interdigital configuration and ii) a concentric interdigital configuration for the nondestructive evaluation of multilayered dielectric structures. A numerical model accounting for sensor geometry, test-piece geometry and real permittivity, and metal electrode thickness has been developed to calculate...
متن کاملZeolite-based Impedimetric Gas Sensor Device in Low-cost Technology for Hydrocarbon Gas Detection
Due to increasing environmental concerns the need for inexpensive selective gas sensors is increasing. This work deals with transferring a novel zeolite-based impedimetric hydrocarbon gas sensor principle, which has been originally manufactured in a costly combination of photolithography, thin-film processes, and thick-film processes to a lowcost technology comprising only thick-film processes ...
متن کامل