Platform switching on wide-diameter external hex implants: a finite element analysis
نویسندگان
چکیده
OBJECTIVES The objective of this work was to use finite element analysis to compare the effect of forces coming to bear on abutments 4.1 or 5.0 mm in diameter connected to a 5.0 mm implant (i.e. with or without platform switching). STUDY DESIGN A 3D CAD model of a 5 x 11.5 mm external hex implant was developed, complete with a connection screw and either of two abutments, one 4.1 and the other 5 mm in diameter, to assess the influence of two loading conditions, i.e. 200 N loaded either axially or off center on the top of the abutment. RESULTS AND CONCLUSIONS In the symmetrically loaded models, greater stresses were transmitted to the bone in the area below the neck of the implant in the case of the wider-diameter abutment. When the narrower abutment was considered, the stress lines remained confined to the metal and were transferred to the bone in a more distal position. When the stresses in the bone where compared under non-symmetrical loading of the larger- and smaller-diameter abutments, the stresses reached lower values in the latter case. These findings indicate that platform switching (i.e. coupling a 4.1 mm abutment with a 5 mm implant) achieves a better, more even distribution of the peri-implant stresses deriving from simulated occlusal loads on the bone margins. Key words:Platform switching, finite element analysis, implant.
منابع مشابه
Clinical Parameters and Crestal Bone Loss in Internal Versus External Hex Implants at One Year after Loading
Introduction: The survival of an implant system is affected by the choice of antirotational design, which can include an external or internal hex. Implant success also is affected by the maintenance of the crestal bone around implants. The aim of present study was to evaluate the crestal bone loss and clinical parameters related to bone loss in patients loaded with an external or internal hex 3...
متن کاملComparative Analysis of Internal and External-Hex Crown Connection Systems - A Finite Element Study
has reduced stress concentrations in the ABSTRACT crown. However, because the torque is transferred through the abutment screw to the abutObjectives: The abutment connection with ment contact, changing the torque has greater the crown is fundamental to the structural effect on this hex system than the masticatory stabil ity of the implant system and to the preforce. Overall the masticatory forc...
متن کاملComparative Evaluation of Osseointegrated Dental Implants Based on Platform-Switching Concept: Influence of Diameter, Length, Thread Shape, and In-Bone Positioning Depth on Stress-Based Performance
This study aimed to investigate the influence of implant design (in terms of diameter, length, and thread shape), in-bone positioning depth, and bone posthealing crestal morphology on load transfer mechanisms of osseointegrated dental implants based on platform-switching concept. In order to perform an effective multiparametric comparative analysis, 11 implants different in dimensions and in th...
متن کاملShort implant to support maxillary restorations: bone stress analysis using regular and switching platform.
PURPOSE The aim of this study was to evaluate stress distribution on peri-implant bone simulating the influence of implants with different lengths on regular and switching platforms in the anterior maxilla by means of three-dimensional finite element analysis. MATERIALS AND METHODS Four mathematical models of a central incisor supported by an external hexagon implant (diameter, 5.0 mm) were c...
متن کاملBiomechanical Effects of Platform Switching in Two Different Implant Systems: A Three-Dimensional Finite Element Analysis
OBJECTIVES The purpose of this study was to determine the influence of platform switching on stress distribution of two different implant systems using three-dimensional (3D) finite element models. MATERIALS AND METHODS Six 3D finite element models were created to replicate two different implant systems with peri-implant bone tissue, in which six different implant-abutment configurations were...
متن کامل