Fracture of cementless femoral stems at the mid-stem junction in modular revision hip arthroplasty systems.

نویسندگان

  • Dror Lakstein
  • Noam Eliaz
  • Ofer Levi
  • David Backstein
  • Yona Kosashvili
  • Oleg Safir
  • Allan E Gross
چکیده

BACKGROUND Mechanical failure of femoral stems at the modular junction of revision hip arthroplasty systems has been reported only infrequently. In the current study, the cause of six stem fractures, which occurred in vivo, was analyzed with use of clinical data and failure analysis. METHODS Six patients with a fracture at the mid-stem junction of a modular revision hip implant were identified in our database of patients who had undergone revision arthroplasty. The characteristics of the patients with a fractured stem were compared with those of 165 patients from the same prospective database who had a modular stem implanted, had at least two years of follow-up, and had not had a fracture of the stem. Failure analysis of three implants (six fracture surfaces) was carried out, with use of microscopic, chemical, and microhardness characterization techniques. RESULTS Patients with a fractured stem had significantly higher body mass indices than patients without a stem fracture. Radiographs demonstrated that these femoral implants lacked adequate osseous support of the junction area of the stem. All stems failed approximately 1 to 2 mm proximal to the body-stem junction, thus indicating the presence of a bending moment. The chemical composition and microhardness matched those of Ti-6Al-4V. Evidence of wear and fatigue were found on the fracture surface. A wear strip was also observed along the circumference of the stem near the junction. CONCLUSIONS We concluded that the stem failure was initiated by a fretting fatigue mechanism and was propagated by a pure bending fatigue mechanism. Risk factors for fractures of the modular junction include excessive body weight and inadequate proximal osseous support because of trochanteric osteotomy, reduced preoperative bone stock, osteolysis, loosening, and/or implant undersizing. Surgeons should consider the use of implants with strengthened junctions when using modular stems in such patients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Total Hip Arthroplasty with Modular Stem for Dysplastic Hips in South Asian Population

Introduction: Optimum component positioning and orientation is required to optimize the functional result during total hip arthroplasty for dysplastic hips. Patients and methods: Sixty-two patients (66 hips) including 33 males and 29 females underwent total hip arthroplasty using modular stem prosthesis at an average age of 40.6 years (range 17 to 49 years). Nineteen hips were clas...

متن کامل

Total Hip Replacement Revision in a Single Brand Small Cementless Stem – Our Experience after the Findings of the National Joint Registry

Background: Cementless total hip replacement is the common THR performed in England, Wales, Northern Irelandand the Isle of Man. The Corail stem is the most popular cementless implant and has a ODEP 10A rating. Review ofits performance in the registry identified an increase rate of revision amongst the smaller stem sizes. However, claritywas not provided on the explanation for this finding. We ...

متن کامل

Outcomes of Dual Modular Cementless Femoral Stems in Revision Hip Arthroplasty

With an increasing number of primary hip replacements being performed every year, the burden of revision hip arthroplasty, for septic and aseptic loosening, recurrent dislocation or periprosthetic fracture, is also increasing. In recent years, different approaches to revising the femoral prosthesis have emerged; including both cemented and cementless techniques. With a stable cement mantle and ...

متن کامل

Mid-term outcome of a modular, cementless, proximally hydroxyapatite-coated, anatomic femoral stem.

PURPOSE To report the mid-term outcome of a modular, cementless, proximally hydroxyapatitecoated, anatomic femoral stem in total hip arthroplasty (THA). METHODS 160 consecutive patients aged 42 to 92 (mean, 70) years underwent 185 cementless THAs for primary osteoarthritis or femoral neck fractures. All procedures were performed by a single surgeon using the same modular, cementless, proximal...

متن کامل

The modern, hybrid total hip arthroplasty for primary osteoarthritis at the Hospital for Special Surgery.

We describe our technique and rationale using hybrid fixation for primary total hip arthroplasty (THA) at the Hospital for Special Surgery. Modern uncemented acetabular components have few screw holes, or no holes, polished inner surfaces, improved locking mechanisms, and maximised thickness and shell-liner conformity. Uncemented sockets can be combined with highly cross-linked polyethylene lin...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. American volume

دوره 93 1  شماره 

صفحات  -

تاریخ انتشار 2011