Automated body-lung-air material map segmentation from pre-clinical MRI images for PET attenuation correction in Tera-Tomo 3D PET reconstruction engine of nanoScan PET/MRI system

نویسندگان

  • Peter Bandi
  • Gabor Jakab
  • Norbert Zsoter
  • Domokos Mathe
  • Gabor Nemeth
  • Kalman Nagy
  • Sandor Hobor
  • Laszlo Papp
چکیده

Department of Software Engineering, Mediso Ltd., Budapest, Hungary The aim of current work was to automatically derive a body-lung-air three-valued material map from MRI images for pre-clinical PET attenuation correction. Our goal was also to measure the effect of attenuation on corresponding pre-clinical PET images. 12 mouse and 9 rat PET/MRI images were acquired by nanoScan PET/MRI (Mediso). Each MRI image was filtered with edge-preserving Non-Local Means filter. Two segmentation stages were determined: the first stage separated the body and the air. The gravitational center of each disjunctive region determined by histogram-derived value ranges of the body and the air were inputs of two independent fuzzy-based affinity map generation steps. The fuzzy maps were merged with simple maximum into an initial binary body-air material map. In the second stage the gravitational center of disjunctive air regions laid inside the body mask were the seed points of the third fuzzy affinity map generation. Three fuzzy maps were merged with maximum operation into a three-valued mask. The three-valued material map was given as input for the Tera-Tomo 3D PET reconstruction engine to serve as the base for attenuation correction. Mouse and rat material map segmentation result were validated by two experienced pre-clinical researchers who manually drawn the ideal mask boundaries. The ratio of voxel misclassification was measured as 2.23% (mouse) and 1.87% (rat). Quantitative ROI measurements shown 1.3 2.7% AC and NAC PET differences. An automated and robust segmentation method was designed to provide a material map for pre-clinical PET attenuation correction. Although current work operated with PET/MRI images, the same method can be applied for nanoScan SPECT/MRI acquisitions. Results show that on the level of real quantitative measurements, attenuation correction of PET makes sense. As the next step of our future work we will investigate the effects of AC on pre-clinical nanoScan SPECT/MRI images. Bandi et al. EJNMMI Physics 2014, 1(Suppl 1):A86 http://www.ejnmmiphys.com/content/1/S1/A86

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Performance Evaluation of Scatter Modeling of the GPU-based “Tera-Tomo” 3D PET Reconstruction

In positron emission tomography (PET), photon scattering inside the body causes significant blurring and quantification error in the reconstructed images. To solve this problem we have developed Monte Carlo (MC) based 3D PET reconstruction algorithms implemented on the Graphics Processing Unit (GPU). Our implementation takes multiple Compton scattering into account without any significant addit...

متن کامل

Comparison of state-of-the-art atlas-based bone segmentation approaches from brain MR images for MR-only radiation planning and PET/MR attenuation correction

Introduction: Magnetic Resonance (MR) imaging has emerged as a valuable tool in radiation treatment (RT) planning as well as Positron Emission Tomography (PET) imaging owing to its superior soft-tissue contrast. Due to the fact that there is no direct transformation from voxel intensity in MR images into electron density, itchr('39')s crucial to generate a pseudo-CT (Computed Tomography) image ...

متن کامل

Tissue classification as a potential approach for attenuation correction in whole-body PET/MRI: evaluation with PET/CT data.

UNLABELLED Attenuation correction (AC) of whole-body PET data in combined PET/MRI tomographs is expected to be a technical challenge. In this study, a potential solution based on a segmented attenuation map is proposed and evaluated in clinical PET/CT cases. METHODS Segmentation of the attenuation map into 4 classes (background, lungs, fat, and soft tissue) was hypothesized to be sufficient f...

متن کامل

A fully automated and reproducible level-set segmentation approach for generation of MR-based attenuation correction map of PET images in the brain employing single STE-MR imaging modality

Quantitative MR Imaging and Spectroscopy Group, Research Center for Cellular and Molecular Imaging, Tehran University of Medical Sciences, Tehran, Iran Generating MR-based attenuation correction map (μ-map) for quantitative reconstruction of PET images still remains a challenge in hybrid PET/MRI systems, mainly because cortical bone structures are indistinguishable from proximal air cavities in...

متن کامل

Emission-based estimation of lung attenuation coefficients for attenuation correction in time-of-flight PET/MR.

In standard segmentation-based MRI-guided attenuation correction (MRAC) of PET data on hybrid PET/MRI systems, the inter/intra-patient variability of linear attenuation coefficients (LACs) is ignored owing to the assignment of a constant LAC to each tissue class. This can lead to PET quantification errors, especially in the lung regions. In this work, we aim to derive continuous and patient-spe...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 1  شماره 

صفحات  -

تاریخ انتشار 2014