Transfer of USP Cholecalciferol Normal-phase HPLC Method to SFC Using the Agilent 1260 Infi nity Hybrid SFC/UHPLC System
نویسنده
چکیده
The Agilent 1260 Infi nity Hybrid SFC/UHPLC System offers a fast, reliable, green, robust, and cost-effective solution by coupling the advantages of HPLC and the benefi ts of supercritical fl uid chromatography (SFC). This Application Note describes the conversion of the United States Pharmacopeia (USP) normal-phase HPLC method for cholecalciferol assay to an SFC method using serially coupled columns. The 1260 Infi nity Hybrid SFC/UHPLC System was used to perform both the HPLC as well as the SFC methods. The results show that the SFC method meets the system suitability criteria, is 2x faster, and achieves 26x lower solvent costs compared to the USP method. Robustness tests on the SFC method demonstrated excellent robustness for routine analysis. Authors Vinayak AK Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India mAU
منابع مشابه
Agilent Application Solution Transfer of a USP method for tolazamide from normal phase HPLC to SFC using the Agilent 1260 Infi nity Hybrid SFC/UHPLC System
Normal phase liquid chromatography methods often have long run times and involve environmentally toxic/costly solvents. Supercritical fl uid chromatography (SFC) methods on the other hand, are faster, inexpensive, and eco-friendly. SFC involves the use of low viscosity supercritical carbon dioxide, which can be operated at fl ow rates up to three times higher than liquid chromatography (LC) wit...
متن کاملAgilent Application Solution Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System
Normal phase liquid chromatography (LC) methods often have long run times and involve environmentally toxic/costly solvents. Supercritical fluid chromatography (SFC) methods on the other hand are faster, inexpensive, and eco-friendly. SFC involves the use of low viscosity supercritical carbon dioxide that can be operated at flow rates up to 3x higher than LC without losing separation efficiency...
متن کاملInjecting variable volumes using the partial loop - fi ll method with the Agilent 1260 Infi nity Analytical SFC System
The Agilent 1260 Infi nity Analytical Supercritical Fluid Chromatography (SFC) System provides LC-like sensitivity and UHPLC power range. It is a pressurized system that has a fi xed-loop autosampler. The injection loops come in different sizes to perform fi xed-loop injections. Injecting a volume smaller than the loop volume is possible using the partial loop-fi ll injection method where the l...
متن کاملDetermination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infi nity Analytical SFC System
This Application Note describes the development of a method for the separation of four stereoisomers, which were derived from a compound with two steric centers. The method was developed using an Agilent 1260 Infi nity Analytical Supercritical Fluid Chromatography (SFC) solution with Agilent ChemStation Method Scouting Wizard software. The developed method was used to compare the ratio of stere...
متن کاملMethod Development to Separate Enantiomers by Supercritical Fluid Chromatography
This Application Note demonstrates the use of the Agilent 1260 Infi nity Analytical SFC Solution for method screening for the separation of enantiomers. To enhance the 1260 Infi nity SFC Solution for method development and screening, a second thermostatted column compartment and two column selection valves were added for column selection, together with a 12-position/13-port valve for solvent se...