Interventional dual-energy imaging-Feasibility of rapid kV-switching on a C-arm CT system.

نویسندگان

  • K Müller
  • S Datta
  • M Ahmad
  • J-H Choi
  • T Moore
  • L Pung
  • C Niebler
  • G E Gold
  • A Maier
  • R Fahrig
چکیده

PURPOSE In the last years, dual-energy CT imaging has shown clinical value, thanks to its ability to differentiate materials based on their atomic number and to exploit different properties of images acquired at two different energies. C-arm CT systems are used to guide procedures in the interventional suite. Until now, there are no commercially available systems that employ dual-energy material decomposition. This paper explores the feasibility of implementing a fast kV-switching technique on a clinically available angiographic system for acquiring dual-energy C-arm CT images. METHODS As an initial proof of concept, a fast kV-switching approach was implemented on an angiographic C-arm system and the peak tube voltage during 3D rotational scans was measured. The tube voltage measurements during fast kV-switching scans were compared to corresponding measurements on kV-constant scans. Additionally, to prove stability of the requested exposure parameters, the accuracy of the delivered tube current and pulse width were also recorded and compared. In a first phantom experiment, the voxel intensity values of the individual tube voltage components of the fast kV-switching scans were compared to their corresponding kV-constant scans. The same phantom was used for a simple material decomposition between different iodine concentrations and pure water using a fast kV-switching protocol of 81 and 125 kV. In the last experiment, the same kV-switching protocol as in the phantom scan was used in an in vivo pig study to demonstrate the clinical feasibility. RESULTS During rapid kV-switching acquisitions, the measured tube voltage of the x-ray tube during fast switching scans has an absolute deviation of 0.23 ± 0.13 kV compared to the measured tube voltage produced during kV-constant acquisitions. The stability of the peak tube voltage over different scan requests was about 0.10 kV for the low and 0.46 for the high energy kV-switching scans and less than 0.1 kV for kV-constant scans, indicating slightly lower stability for kV-switching scans. The tube current resulted in a relative deviation of -1.6% for the low and 6.6% overestimation for the high tube voltage of the kV-switching scans compared to the kV-constant scans. The pulse width showed no deviation for the longer pulse width and only minor deviations (0.02 ± 0.02 ms) for the shorter pulse widths compared to the kV-constant scans. The phantom experiment using different iodine concentrations showed an accurate correlation (R2 > 0.99) between the extracted intensity values in the kV-switching and kV-constant reconstructed volumes, and allows for an automatic differentiation between contrast concentration down to 10% (350 mg/ml iodine) and pure water under low-noise conditions. Preliminary results of iodine and soft tissue separation showed also promising results in the first in vivo pig study. CONCLUSIONS The feasibility of dual-energy imaging using a fast kV-switching method on an angiographic C-arm CT system was investigated. Direct measurements of beam quality in the x-ray field demonstrate the stability of the kV-switching method. Phantom and in vivo experiments showed that images did not deviate from those of corresponding kV-constant scans. All performed experiments confirmed the capability of performing fast kV-switching scans on a clinically available C-arm CT system. More complex material decomposition tasks and postprocessing steps will be part of future investigations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Interventional Dual - Energy Imaging 1 - Feasibility of rapid kV - switching on a C - Arm CT System

Purpose: In the last years, dual-energy CT imaging has shown clinical value thanks to its ability to differentiate materials based on their atomic number and to exploit different properties of images acquired at two different energies. C-arm CT systems are used to guide procedures in the interventional suite. Up to now, there are no commercially available systems that employ dual-energy materia...

متن کامل

Dual-energy CT in the Chest

Dual-energy CT (DECT) is accomplished by near simultaneous imaging at two energy levels, typically 80 kV and 140 kV, using scanners that either employ rapid kV switching or have two orthogonal sets of X-ray tunes and detectors mounted in the gantry. Attenuation properties differ at the two energy levels, which allow the DECT to differentiate and quantify certain substances such as water, calciu...

متن کامل

Dual-energy CT revisited with multidetector CT: review of principles and clinical applications.

Although dual-energy CT (DECT) was first conceived in the 1970s, it was not widely used for CT indications. Recently, the simultaneous acquisition of volumetric dual-energy data has been introduced using multidetector CT (MDCT) with two X-ray tubes and rapid kVp switching (gemstone spectral imaging). Two major advantages of DECT are material decomposition by acquiring two image series with diff...

متن کامل

Images in cardiovascular medicine. Myocardial ischemia diagnosed by dual-energy computed tomography: correlation with single-photon emission computed tomography.

For decades, dual-energy imaging has been used for tissue differentiation with several x-ray–based imaging modalities, exploiting the fact that the tissues in the human body show different absorption characteristics when penetrated with different x-ray spectra, spectra that are typically generated by different kV settings of the x-ray tube. Recently, dual-source computed tomography (CT) with 2 ...

متن کامل

Estimation of fetal absorbed dose from low-dose attenuation-correction CT in PET / CT imaging by using the Body Builder Phantom

One of the methods of studying the physiology and metabolism of important tissues such as the heart, brain and cancer tumors is the use of PET/CT System. A small number of patients are pregnant women who undergo a PET/CT scan due to lack of knowledge about pregnancy or due to dire clinical need. The final dose received by the fetus is based on three factors: the absorbed dose of the fetal tissu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Medical physics

دوره 43 10  شماره 

صفحات  -

تاریخ انتشار 2016