Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents
نویسندگان
چکیده
PURPOSE To validate the Dinamap ProCare 200 blood pressure (BP) monitor against a mercury sphygmomanometer in children 7 to 18 years old in accordance with the 2010 International Protocol of European Society of Hypertension (ESH-IP2) and the British Hypertension Society (BHS) protocol. METHODS Forty-five children were recruited for the study. A validation procedure was performed following the protocol based on the ESH-IP2 and BHS protocols for children and adolescents. Each subject underwent 7 sequential BP measurements alternatively with a mercury sphygmomanometer and the test device by trained nurses. The results were analyzed according to the validation criteria of ESH-IP2. RESULTS The mean (±SD) difference in the absolute BP values between test device and mercury sphygmomanometer readings was 1.85±1.65 mmHg for systolic BP (SBP) and 4.41±3.53 mmHg for diastolic BP (DBP). These results fulfilled the Association for the Advancement of Medical Instrumentation criterion of a mean±SD below 5±8 mmHg for both SBP and DBP. The percentages of test device-observer mercury sphygmomanometer BP differences within 5, 10, and 15 mmHg were 96%, 100%, and 100% for SBP, and 69%, 92%, and 100% for DBP, respectively, in the part 1 analysis; both SBP and DBP passed the part 1 criteria. In the part 2 analysis, SBP passed the criteria but DBP failed. CONCLUSION Although the Dinamap ProCare 200 BP monitor failed an adapted ESH-IP2, SBP passed. When comparing BP readings measured by oscillometers and mercury sphygmomanometers, one has to consider the differences between them, particularly in DBP, because DBP can be underestimated.
منابع مشابه
Automated Versus Manual Blood Pressure Measurement: A Randomized Crossover Trial in the Emergency Department of a Tertiary Care Hospital in Karachi, Pakistan: Are Third World Countries Ready for the Change?
BACKGROUND Hypertension has proven to be a strong liability with 13.5% of all mortality worldwide being attributed to elevated blood pressures in 2001. An accurate blood pressure measurement lies at the crux of an appropriate diagnosis. Despite the mercury sphygmomanometer being the gold standard, the ongoing deliberation as to whether mercury sphygmomanometers should be replaced with the autom...
متن کاملComparison of blood pressure measurements on the upper and lower extremities during spinal anesthesia
Introduction: During spinal anesthesia, measurement of blood pressure is often obtained by an indirect method using an inflatable cuff on the upper arm. The relationship between the blood pressure measurements at the two sites in patients undergoing spinal anesthesia is unclear. This study was designed to evaluate the difference observed in systolic blood pressure (SBP) taken from the upp...
متن کاملSources of variability in blood pressure measurement using the Dinamap PRO 100 automated oscillometric device.
The Dinamap automated oscillometric device (GE Medical Systems Information Technologies, Inc., Milwaukee, Wisconsin) for blood pressure measurement is widely employed in epidemiologic studies because it is easy to use and because it eliminates observer variability. In this study, the authors assessed the variability in observed blood pressures associated with use of the Dinamap monitor and esti...
متن کاملSkip patterns in DINAMAP-measured blood pressure in 3 epidemiological studies.
Blood pressure measured by the oscillometric, automated device DINAMAP in 3 large population-based studies sponsored by the National Heart, Lung, and Blood Institute (The Atherosclerosis Risk in Communities Study, The Family Heart Study, and the Hypertension Genetic Epidemiology Network Study) were reviewed to determine an apparent skip pattern in the measurement values. Across the 3 studies, 2...
متن کاملValidation of the YuWell YE690A upper-arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010.
OBJECTIVE The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure monitor YuWell YE690A for blood pressure measurement according to the International Protocol of the European Society of Hypertension revision 2010. PARTICIPANTS AND METHODS Systolic and diastolic blood pressures were measured sequentially in 33 adult Chinese (12 women, 44.2 years...
متن کامل