Benchmarking the gamma pass score using ArcCHECK for routine dosimetric QA of VMAT plans
نویسندگان
چکیده
A minimum expected gamma (γ) pass rate for VMAT plan verification using ArcCHECK was established based on the RTTQA, TG119 test cases and 10 clinical plans with varying levels of complexity. The impact of the ‘Measurement Uncertainty’ parameter as available in the ArcCHECK software on γ pass rate was studied for both global and local γ analysis. Our results show that excluding measurement uncertainty adds tighter tolerance in local γ comparison. From the verification of our benchmark cases we established minimum expected γ pass rates of 85% and 88% for 2%/2mm global and 3%/3mm local tolerance criteria.
منابع مشابه
Statistical process control analysis for patient-specific IMRT and VMAT QA
This work applied statistical process control to establish the control limits of the % gamma pass of patient-specific intensity modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) quality assurance (QA), and to evaluate the efficiency of the QA process by using the process capability index (Cpml). A total of 278 IMRT QA plans in nasopharyngeal carcinoma were measured with ...
متن کاملThe quality assurance of volumetric modulated arc therapy (VMAT) plans for early stage prostate cancer: a technical note
As radiation therapy transitions from intensity modulated radiation therapy (IMRT) to volumetric modulated arc therapy (VMAT) it is important to consider the quality assurance (QA) of VMAT plans in light of what has previously been learned and developed in IMRT QA. This technical note assesses if IMRT based plan QA software, which has reduced the need in IMRT for phantom dose measurements on th...
متن کاملEvaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT
Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...
متن کاملEvaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric‐modulated arc therapy (VMAT)
The purpose of this study is to evaluate the sensitivities of 3D diode arrays to setup error for patient-specific quality assurance (QA) of volumetric-modulated arc therapy (VMAT). Translational setup errors of ± 1, ± 2, and ± 3 mm in the RL, SI, and AP directions and rotational setup errors of ± 1° and ± 2° in the pitch, roll, and yaw directions were set up in two phantom systems, ArcCHECK and...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره شماره
صفحات -
تاریخ انتشار 2014