Comparison of the fracture resistance of dental implants with different abutment taper angles.

نویسندگان

  • Kun Wang
  • Jianping Geng
  • David Jones
  • Wei Xu
چکیده

To investigate the effects of abutment taper angles on the fracture strength of dental implants with TIS (taper integrated screwed-in) connection. Thirty prototype cylindrical titanium alloy 5.0mm-diameter dental implants with different TIS-connection designs were divided into six groups and tested for their fracture strength, using a universal testing machine. These groups consisted of combinations of 3.5 and 4.0 mm abutment diameter, each with taper angles of 6°, 8° or 10°. 3-Dimensional finite element analysis (FEA) was also used to analyze stress states at implant-abutment connection areas. In general, the mechanical tests found an increasing trend of implant fracture forces as the taper angle enlarged. When the abutment diameter was 3.5 mm, the mean fracture forces for 8° and 10° taper groups were 1638.9 N ± 20.3 and 1577.1 N ± 103.2, respectively, both larger than that for the 6° taper group of 1475.0 N ± 24.4, with the largest increasing rate of 11.1%. Furthermore, the difference between 8° and 6° taper groups was significant, based on Tamhane's multiple comparison test (P<0.05). In 4.0 mm-diameter abutment groups, as the taper angle was enlarged from 6° to 8° and 10°, the mean fracture value was increased from 1066.7 N ± 56.1 to 1241.4 N ± 6.4 and 1419.3 N ± 20.0, with the largest increasing rate of 33.1%, and the differences among the three groups were significant (P<0.05). The FEA results showed that stress values varied in implants with different abutment taper angles and supported the findings of the static tests. In conclusion, increases of the abutment taper angle could significantly increase implant fracture resistance in most cases established in the study, which is due to the increased implant wall thickness in the connection part resulting from the taper angle enlargement. The increasing effects were notable when a thin implant wall was present to accommodate wide abutments.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the Effect of Two Different Abutment Diameters on Detorque Value Using Cyclic Loading and Thermocycling

Background and Aim: Screw loosening is a common problem with both screw-retained and cemented implant restorations. It is assumed that the abutment diameter affects detorque value and screw loosening. We aimed to determine the effect of two different abutment diameters on detorque value using cyclic loading and thermocycling. Materials and Methods: This in-vitro experimental study was conducte...

متن کامل

Abutment fracture and the technique Used to Remove the fracture portion of the Abutmet: Case Report

Abstract This clinical report presents a case of abutment fracture, due to occlusal overload, occurred with an implant-supported restoration on right mandibular first molar of a female patient. After 6 months of prosthetic reconstruction implants, the patient returned to the clinic due to the loss of implant crown. The fractured fragment of the abutment with the screw protec...

متن کامل

مقایسه مقاومت شکست و نوع شکست اباتمنت‌های زیرکونیایی مختلف در مقایسه با اباتمنت‌های تیتانیومی

Background and Aims: One of the decisive factors on the final esthetic results of implant treatment is the abutment material. Ceramic abutments are nowadays very popular in this regard. The purpose of this study was to compare the fracture resistance of three zirconia abutments with titanium abutment. Materials and Methods: Four groups of abutments (five specimens in each) were prepared. The a...

متن کامل

Evaluating the impact of length and thread pitch on the stress distribution in dental implants and surrounding bone using finite element method

 longevity of osseointegrated implants are intensely influenced by biomechanical factors. Control of these factors prevents mechanical complications, which include fracture of screws, components, or materials veneering the framework. In this study, the impact of length and threads pitch of dental implants on the stress distribution and maximum Von Mises stress in implant-abutment complex and ja...

متن کامل

Mechanical testing of implant-supported anterior crowns with different implant/abutment connections.

PURPOSE This study evaluated the reliability and failure modes of anterior implants with internal-hexagon (IH), external-hexagon (EH), or Morse taper (MT) implant-abutment interface designs. The postulated hypothesis was that the different implant-abutment connections would result in different reliability and failure modes when subjected to step-stress accelerated life testing (SSALT) in water....

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Materials science & engineering. C, Materials for biological applications

دوره 63  شماره 

صفحات  -

تاریخ انتشار 2016