False-positive results with an alcometer device.
نویسندگان
چکیده
We write to report false-positive results with an alcometer device (Lion S-D2 brand) when it is used in the vicinity of recently applied alcohol disinfectant hand rub. The Orchard Clinic is a medium secure forensic unit set within the grounds of the Royal Edinburgh Hospital. The use of an alcometer is essential in the treatment of Scottish government 'restricted' patients or if there is suspicion or evidence of alcohol misuse. Indeed, all staff receive training on the use of the alcometer during their induction period. Recently a patient with 'restricted' status returned from leave outwith the clinic. On the patient's return they were breathalysed. The alcometer gave a reading of 0.04 mg/lBrAC – equivalent to just under one unit of alcohol. The patient denied consuming alcohol. Staff had used an alcohol-based hand disinfectant lotion prior to carrying out the test. This was consistent with hand hygiene guidelines following the H1N1 flu pandemic. We later reproduced the positive alcometer results after using alcohol hand lotion in the standard way. The reading rose from 0.01 mg/lBrAC (background reading) to 0.1 mg/lBrAC – equivalent to two units of alcohol. The alcohol hand rub used within NHS Lothian is up to 80% alcohol, which is one of the highest on the market. 1 It is therefore likely there is much alcohol in the vapour around where it is used. When the individual inhales prior to the test, some alcohol vapour is also taken in which is exhaled back into the instrument as part of their sample. False-positive alcometer results may have serious implications for the treatment of forensic patients. No other report of alcohol hand lotion raising alcometer readings was found in the literature. It had not been reported as an issue with local colleagues in the Alcohol Problem Service or with Lothian and Borders Police. However, it is recognised that the police may carry out the test in a more open environment than a ward treatment room. In addition to reporting this finding to the manufacturer, the local procedure for operating the alcometer has now been amended to ensure that alcohol hand lotion: (a) is not used by any staff involved in taking an alcometer reading at least 5 min before taking the test; and (b) is not used in the same room at least 5 min before a test is carried out. Staff are encouraged to use soap and water to clean …
منابع مشابه
آلودگی باکتریال کشتهای خون مرکز طبی کودکان تهران - فروردین تا خرداد 1383
Background: Blood culture is the criterion standard for identifying children with bacteremia. However, elevated false-positive rates are common and are associated with substantial health care costs. The aims of this prospective study were to: 1) determine the rate of blood culture contamination 2) determine variety and frequency of contaminant bacteria 3) compare the duration of hospital stay a...
متن کاملThe Development of a Passive False Twister Mechanism in Handling Low Strength Cotton Slivers on High Draft Spinning Machine
A passive false twist unit (spiral) has been developed to assist with the handling lowstrength slivers on a high speed-spinning machine with a high-speed feed. In the first trial, a falsetwist simulator device was constructed to determine whether the passive false twist unit can be usedon high speed feeding with different can distance from the feeding device. In the second trial, theeffects of ...
متن کاملThe prevalence of adverse obstetric outcomes follow of an abnormal maternal serum AFP and NTDs screen positive group
Introduction: Prenatal maternal serum AFP screening tests have some inherent false positive and false negative results. For NTDs Screening test (FTS) false positive rate (FPR) is about 1% and false negative rate (FNR) is 20%. A false positive result encourages by Anomaly scan and invasive diagnostic procedures (that poses lots of stress on women, too), while a false negative result lead to a N...
متن کاملMultiple non-malignant iodine uptake in a 131I whole body scan of a patient with papillary thyroid carcinoma: Importance of SPECT/CT
After thyroidectomy in papillary thyroid carcinoma (PTC) patients, whole body radioiodine scan (WBIS) is necessary for localization of any abnormal radioiodine avid lesions. However false-positive radioiodine uptake in WBIS can be problematic. We report a 32 years old female patient with PTC. After thyroidectomy and radioiodine therapy WBIS was performed. In her post therapy WBIS, there were zo...
متن کاملCan chest trauma patients provide breath sample with Lion SD-400 Alcometer?
Various investigators have addressed the minimum lung function required to activate breathalyzers, and the impact of comorbid respiratory illness. We postulated that subjects with significant chest trauma may have difficulty in providing an adequate breathalyzer sample. A prospective self-controlled study of 20 patients who underwent thoracotomy was conducted between August 2005 and December 20...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The British journal of psychiatry : the journal of mental science
دوره 197 1 شماره
صفحات -
تاریخ انتشار 2010