Coil thickness and packing of cerebral aneurysms: a comparative study of two types of coils.

نویسندگان

  • Marjan J Slob
  • Willem Jan van Rooij
  • Menno Sluzewski
چکیده

BACKGROUND AND PURPOSE In coiling cerebral aneurysms, high packing prevents reopening over time. The purpose of this study was to compare packing of cerebral aneurysms treated with two types of coils with different wire thickness and different shapes. METHODS Packing, defined as the ratio between the volume of inserted coils and the volume of an aneurysm, was calculated for 144 cerebral aneurysms treated in 130 patients. Seventy-two aneurysms were treated with predominantly helical-shaped coils of 0.010-inch-diameter wire, and 72 aneurysms were treated with predominantly complex-shaped coils of 0.012-inch-diameter wire. Aneurysm volume was assessed from three-dimensional angiography. Aneurysm packing, inserted coil lengths, and numbers of coils were compared for both types of coils. RESULTS Mean packing was significantly higher (absolute value, 6.4%; relative value, 26.6%; P < .0001) in aneurysms coiled with 0.012-inch-diameter coils than aneurysms coiled with 0.010-inch-diameter coils. Inserted coil length per cubic millimeter of aneurysmal volume was equal for both types of coils. CONCLUSION Use of complex-shaped coils with a wire diameter of 0.012 inch to coil cerebral aneurysms results in significantly better packing than is achieved with helical coils of 0.010-inch-diameter wire.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Practical Feasibility and Packing Density of Endovascular Coiling Using Target® Nano™ Coils in Small Cerebral Aneurysms

OBJECTIVE Based on the use of Nano™ coils, we retrospectively compared the proportion of the coils (≤ 1.5 mm) and packing density in two patient groups with small cerebral aneurysms (< 4 mm diameter) who were treated with or without Nano™ coils. MATERIALS AND METHODS Between January 2012 and November 2013, in 548 cerebral aneurysms treated by endovascular coiling, 143 patients with 148 small ...

متن کامل

Packing performance of helical Guglielmi detachable coil (GDC) 18 in intracranial aneurysms: a comparison with helical GDC 10 coils and complex Trufill/Orbit coils.

BACKGROUND AND PURPOSE The purpose of this study was to retrospectively compare the packing performance of helical Guglielmi detachable coil (GDC) 18 (thickness, 0.0135-0.015 inch) with the packing performance of both helical GDC 10 (thickness, 0.010 inch) and complex Trufill/Orbit coils (thickness, 0.012 inch). MATERIALS AND METHODS From our data base, we selected aneurysms that were exclusi...

متن کامل

Matrix and bioabsorbable polymeric coils accelerate healing of intracranial aneurysms: long-term experimental study.

BACKGROUND AND PURPOSE Acceleration of intra-aneurysmal clot organization and fibrosis may be a solution to preventing aneurysm recanalization after endovascular treatment. The purpose of this study was to evaluate the short-term efficacy and long-term safety of the new Matrix coil system. METHODS Matrix coils consist of thin platinum coils covered with a bioabsorbable, polymeric material (po...

متن کامل

Mathematic analysis of incremental packing density with detachable coils: does that last coil matter much?

Higher packing attenuation of coils in cerebral aneurysms is associated with a decreased recurrence rate. However, geometric relationships suggest that an additional coil may have very little effect on packing attenuation as aneurysm size increases. We mathematically evaluated the relationship between aneurysm size and incremental packing attenuation for coils currently available.

متن کامل

Endovascular Embolization of Intracranial Aneurysms Using Bare Platinum Axium™ Detachable Coils: Immediate and Short-Term Follow-up Results from a Multicenter Registry

PURPOSE Axium™ coils were developed to improve the durability of coil-embolized cerebral aneurysms by increasing packing density. The purpose of this prospective multicenter registry was to evaluate the safety and durability of Axium™ coils. MATERIALS AND METHODS One hundred twenty-six patients with 135 aneurysms of ≤ 15 mm in size underwent coil embolization using bare platinum coils, with A...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 26 4  شماره 

صفحات  -

تاریخ انتشار 2005