Evaluation of six inhalable aerosol samplers.
نویسندگان
چکیده
Six inhalable aerosol samplers were evaluated experimentally as area samplers using monodisperse solid particles with aerodynamic diameters ranging from 5 to 68 microm. Sampler performance and inside particle loss at two test wind speeds (0.55 and 1.1 m/sec) and three wind orientations (0, 90, and 180 degrees) were investigated. The six inhalable aerosol samplers tested were a RespiCon, an Institute of Occupational Medicine (IOM), a seven-hole, a conical inhalable sampler, a prototype button sampler, and a closed-face 37-mm cassette. The area sampling performance of the RespiCon sampler matched the inhalable convention fairly well. The sampling performances of the other five samplers depended on wind speed, wind direction, and particle size, and they may not be appropriate for area sampling if the wind speeds are greater than 0.5 m/sec.
منابع مشابه
Research Papers Involving the Tsi Respicontm Sampler
Six inhalable aerosol samplers were evaluated experimentally as area samplers using monodisperse solid particles with aerodynamic diameters ranging from 5 to 68 mm. Sampler performance and inside particle loss at two test wind speeds (0.55 and 1.1 m/sec) and three wind orientations (0, 90, and 180 degrees) were investigated. The six inhalable aerosol samplers tested were a RESPICONTM sampler, a...
متن کاملResearch Papers Involving the TSI RespiCon Sampler - Application Note ITI-080
Six inhalable aerosol samplers were evaluated experimentally as area samplers using monodisperse solid particles with aerodynamic diameters ranging from 5 to 68 mm. Sampler performance and inside particle loss at two test wind speeds (0.55 and 1.1 m/sec) and three wind orientations (0, 90, and 180 degrees) were investigated. The six inhalable aerosol samplers tested were a RESPICONTM sampler, a...
متن کاملLaboratory study of selected personal inhalable aerosol samplers.
Assessment of inhalable dust exposure requires reliable sampling methods in order to measure airborne inhalable particles' concentrations. Many inhalable aerosol samplers can be used but their performances widely vary and remain unknown in some cases. The sampling performance of inhalable samplers is strongly dependent on particle size and ambient air velocity. Five inhalable aerosol samplers h...
متن کاملPerformance of personal inhalable aerosol samplers in very slowly moving air when facing the aerosol source.
While personal aerosol samplers have been characterized primarily based on wind tunnel tests conducted at relatively high wind speeds, modern indoor occupational environments are usually represented by very slow moving air. Recent surveys suggest that elevated levels of occupational exposure to inhalable airborne particles are typically observed when the worker, operating in the vicinity of the...
متن کاملA collaborative European study of personal inhalable aerosol sampler performance.
Following the adoption of new international sampling conventions for inhalable, thoracic and respirable aerosol fractions, a working group of Comité Européen de Normalisation (CEN) drafted a standard for the performance of workplace aerosol sampling instruments. The present study was set up to verify the experimental, statistical and mathematical procedures recommended in the draft performance ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- AIHAJ : a journal for the science of occupational and environmental health and safety
دوره 61 4 شماره
صفحات -