Calibration and validation of EchoMRI™ whole body composition analysis based on chemical analysis of piglets, in comparison with the same for DXA.

نویسندگان

  • Israel Kovner
  • Gersh Z Taicher
  • Alva D Mitchell
چکیده

A study was conducted to appraise a new EchoMRI™ device for body composition analysis (BCA) of infants and to compare it with dual energy X-ray absorptiometry (DXA), using chemical analysis as a reference method.The calibration part of the study included cross-validation comparisons between EchoMRI™ measurements of awake, anesthetized and dead piglets of the calibration set. It also included comparison of two different approaches to refining the calibration of EchoMRI™, by low- or by high-dimensional linear regressions. Only the low-dimensional approach was applied to DXA.The validation part yielded EchoMRI™ accuracy of 27 g and 70 g for fat and total water, respectively, on piglets scanned while anesthetized, as compared with 24 g and 57 g, respectively, for DXA.EchoMRI™ precision was found to be 4 g and 7 g for fat and total water, respectively, for anesthetized piglets, as compared to 16 g and 14 g, respectively, for DXA. The differences between fat measurements of awake, anesthetized and dead piglets can be statistically significant, but are comparable in magnitude to random errors.To summarize: Characterization of random errors by CV, especially that of fat, is not suitable for BCA, whereas absolute errors or errors relative to total body weight can be applicable. Low- and high-dimensional regressions offer nearly the same accuracy improvements. Improved DXA and EchoMRI™ offer nearly the same accuracy, within 1% of weight in fat, while EchoMRI™ has better precision, within 0.2 % of weight in fat for anesthetized and dead piglets as compared to DXA's 0.5-0.6%.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a new body composition phantom for quality control and cross-calibration of DXA devices.

This study evaluated a new body composition phantom and its use for quality control and cross-calibration of dual-energy X-ray absorptiometry (DXA) instruments for measurements of body composition. We imaged the variable composition phantom (Lunar, Madison, WI) on eight different DXA devices. Deviations of up to 7% fat were observed when we compared the percent fat values measured by the differ...

متن کامل

Fat requirement of Bester in winter feeding; effect of fat levels on growth indices and body chemical composition

This study aimed to determine the fat requirement and effect of dietary fat levels on growth and body composition indices of Bester (Huso huso ♀ × Acipenser ruthenus ♂) juveniles( 200.3 ± 2 g) in winter feeding. Fish were fed by 5 isonitrogenous diets (40% crude protein) with different fat levels including 10, 13, 16, 19 and 22%, 3 times daily according to the satiation for 50 days. The results...

متن کامل

Comparison between live food and artificial diet on survival rate, growth and body chemical composition of Oncorhynchus mykiss larvae

  This study was conducted to evaluate the efficacy of live food (Artemia urmiana) and commercial diet in rearing of Oncorhynchus mykiss larvae. Experiment started when larvae were 0.12g in weight. Triplicate groups of fish were offered one of four treatments: (1) a commercial starter food (2) live Artemia urmiana nauplii (for 3 days) (3) live Artemia urmiana nauplii (for 7 days) and (4) combin...

متن کامل

Invited Review Use of dual-energy x-ray absorptiometry in non-ruminant nutrition research

Precise body composition measurements are essential in animal nutrition studies because the impact of treatments is evaluated based on changes in body weight and composition. Various indirect techniques for animal compositional evaluation have been developed and evaluated for applicability in animal nutrition studies. A fast, accurate, minimally invasive method that requires little input is con...

متن کامل

Comparison of Body Composition Indices of Chemical Warfare Veterans with Chronic Pulmonary Obstruction in Patients with Low and Moderate to High Disease Severity

Background and Aim: More than 100,000 Iranian veterans suffer from chemical damage caused by mustard gas. The aim of this study was to determine the body composition status of chemical warfare veterans with chronic obstructive pulmonary disease (COPD). Methods: This is a cross-sectional study. 97 chemical warfare veterans with COPD in Karaj, Iran were included in the study. The Patients were di...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • International journal of body composition research

دوره 8 1  شماره 

صفحات  -

تاریخ انتشار 2010