Shapability, memory, and luminal changes in microcatheters after steam shaping: a comparison of 11 different microcatheters.
نویسندگان
چکیده
PURPOSE The purpose of this study was to compare the characteristics of shaped microcatheters, including shapability, durability, and luminal changes. MATERIALS AND METHODS Eleven brands of steam-shaped microcatheters and one brand of preshaped microcatheter were evaluated. There were 2 nonreinforced and 10 reinforced devices supported by coils. For evaluation of shapability, the tip angle of 6 samples of each brand were measured after steam-shaping for 20 seconds with a shaping mandrel bent at a 90 degrees or 150 degrees angle. The ability to maintain the shaped angle after guidewire insertion stress (durability) was compared by calculation of the change in the tip angle by using 3 samples of each brand. Luminal change after steam shaping was evaluated by calculation of narrowing rate of the smallest diameter and observation of the surface morphology of the mold of each catheter lumen by using a silicone polymer by means of a fluorescent projection method. RESULTS The nonreinforced microcatheters and the fiber-braided microcatheter showed higher shapability than the others. The degree of distal microcatheter straightening with the microguidewire insertion was less pronounced in the preshaped microcatheter and the fiber-braided microcatheter. Spontaneous recovery to the initial tip angle 5 minutes after the guidewire procedure was observed in 10 brands to various degrees (87%-98%). Irregular luminal surface morphology at the angled portion was found in 6 reinforced brands. One nonreinforced catheter and the fiber-braided catheter showed high narrowing rates >6%. CONCLUSION There are differences in shapabilty, durability, and luminal changes of steam shaping in 12 brands of microcatheters. These characteristics could be important factors in catheter choice for endovascular procedures.
منابع مشابه
Distal-tip shape-consistency testing of steam-shaped microcatheters suitable for cerebral aneurysm coil placement.
BACKGROUND AND PURPOSE Although the selection of microcatheter for endovascular aneurysmal treatment is one important factor in patient outcome, the use of steam shaping for achieving safe entry and stability during coil placement has not, to our knowledge, been systematically evaluated. The goal of this study was to compare the durability of distal microcatheter steam shaping in five different...
متن کاملMarker alignment for Guglielmi detachable coil embolization: practical considerations.
BACKGROUND AND PURPOSE Aneurysm embolization is not without risk: numerous technical aspects are considered before, during, and after the procedure. The purpose of this study was to show the position of the detachment zone of a Guglielmi detachable coil (GDC) with respect to the catheter tip for various microcatheters and marker alignments. METHODS Six types of commonly used microcatheters we...
متن کاملPerformance characteristics of microcatheter systems in a standardized tortuous pathway.
BACKGROUND AND PURPOSE Published reports of controlled experiments designed to evaluate the performance of over-the-wire microcatheter systems are rare and have often been based on subjective impressions from small clinical series. This investigation was designed to compare the load forces required to propel state-of-the-art, hydrophilically coated microcatheters from each of four manufacturers...
متن کاملSafety of retained microcatheters: an evaluation of radiofrequency heating in endovascular microcatheters with nitinol, tungsten, and polyetheretherketone braiding at 1.5 T and 3 T.
BACKGROUND The use of ethylene-vinyl alcohol copolymer for liquid embolization of cranial vascular lesions has resulted in microcatheter fragments entrapped in patients following endovascular procedures. Undergoing subsequent diagnostic MRI examinations poses a safety concern due to the possibility of radiofrequency heating of the metallic braid incorporated into the microcatheter. Heating of n...
متن کاملThe Interference Phenomenon of Microcatheters in the Jailing Treatment for Internal Carotid Artery Side Wall Aneurysms with an Open Cell Stent System
OBJECTIVE Excelsior XT-27 (Stryker Neurovascular, Fremont, CA, USA) or Rebar 27 (eV3 Covidien, Irvine, CA, USA) microcatheters have recently been used to overcome the limitations of the Renegade Hi-Flo microcatheter such as interference between two microcatheters (one for stent delivery and the other for cerebral aneurysm coiling) during the jailing technique. We evaluated differences and influ...