Anterior segment biometry with the Pentacam: comprehensive assessment of repeatability of automated measurements.

نویسندگان

  • Hema Shankar
  • Deepa Taranath
  • Chandramalar T Santhirathelagan
  • Konrad Pesudovs
چکیده

PURPOSE To comprehensively assess the reliability of automated Pentacam (Oculus, Inc.) measurements. SETTING Flinders Eye Centre, Flinders Medical Centre, Bedford Park, South Australia, Australia. METHODS Both eyes of 35 normal volunteers were tested twice on the same day by 2 different observers. All automated values were recorded, and manual analysis of topographic maps was performed only to overrule variance in corneal thickness due to pupil decentration altering the central reference point. Repeatability was determined with Bland-Altman limits of agreement and reported as the coefficient of repeatability (COR = +/-1.96 standard deviation of differences). Relative repeatability (RR) was calculated as a percentage of the ratio of COR to the mean. RESULTS Overall, repeatability was good. Corneal curvature, reported in diopters, showed good repeatability anteriorly (simulated keratometry mean COR+/-0.28D; RR=0.64%) and posteriorly (COR+/-011D; RR=1.85%). Peripheral corneal curvature was more reliable when calculated by the sagittal (axial) method (RR=1.57%) than by the tangential (meridional) method (RR=2.38%). Keratometric power deviation was less reliable (RR=16.39%). Anterior chamber measurements showed good reliability (RR=3.07%-5.68%) except for anterior chamber angle (RR=14.41%). Pupil diameter showed poor reliability (RR=25.77%). Central corneal thickness was comparable at pupil center and corneal vertex, but peripheral repeatability was much better when centered on the corneal vertex (COR+/-16.00microm; RR=2.56%) than at pupil center (COR+/-26.28microm; RR=4.23%). CONCLUSIONS Pentacam corneal curvature and anterior chamber parameters were highly repeatable, but pupil measurements had poor repeatability. Peripheral pachymetry readings were affected by pupil decentration and required manual analysis using the corneal vertex as the point of reference to achieve good repeatability.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

مقایسه‌ی کراتومتری پنتاکم وکراتومتری IOL Master در محاسبه‌ی لنز داخل چشمی افراد با قرنیه‌ی طبیعی

Background and Objective: Proper method and machine for corneal evaluation is an important factor in many anterior segment interventions. This study was performed to compare the mean keratometry (K) readings obtained with an IOL Master and Pentacam for intraocular lens power calculation in normal subjects with no history of refractive surgery. Materials and Methods: Mean K values were obtained ...

متن کامل

Quantitative assessment of anterior chamber volume using slit-lamp OCT and Pentacam.

PURPOSE To quantitatively assess and compare the anterior chamber volume (ACV) by different anterior segment imaging techniques. METHODS Forty healthy eyes of 40 patients were recruited and three consecutive measurements of ACV were determined prospectively utilizing SL-OCT and Pentacam. Statistical significance of interdevice differences between measurements was evaluated by unpaired t test ...

متن کامل

Validity and repeatability of anterior chamber depth measurements with Pentacam and Orbscan.

PURPOSE The purpose of this study was to determine the validity and repeatability of anterior chamber depth (ACD) measurements obtained with a novel rotating Scheimpflug camera (Pentacam; Oculus Optikgeräte GmbH, Wetzlar, Germany) to scanning slit topography (Orbscan; Bausch & Lomb, Rochester, NY). METHODS ACD in 60 healthy eyes was measured twice by two independent observers with each modali...

متن کامل

A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices

PURPOSE To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices. METHODS Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measurements with the RC-5000 (Tomey Corp., Japan), KR-8000 (Topcon, Japan), IOLMaster (Carl Zeiss Med...

متن کامل

Efficacy of Orbscan II and Pentacam topographers by a repeatability analysis when assessing elevation maps in candidates to refractive surgery.

INTRODUCTION Anterior and posterior corneal elevations are measurements used to detect keratoconus suspects. PURPOSE To determine the efficacy of Orbscan II and Pentacam when assessing their elevation maps. MATERIALS AND METHODS The efficacy of the Orbscan II and Pentacam measuring the anterior and posterior corneal elevations were evaluated in a sample of 68 eyes. The concordance between t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of cataract and refractive surgery

دوره 34 1  شماره 

صفحات  -

تاریخ انتشار 2008