The step‐and‐shoot IMRT overshooting phenomenon: a novel method to mitigate patient overdosage
نویسندگان
چکیده
The goal of this work is to evaluate the dosimetric impact of an overshooting phenomenon in step-and-shoot IMRT delivery, and to demonstrate a novel method to mitigate the issue. Five pelvis IMRT patients treated on Varian 2100C EX linacs with larger than +4.5% phantom ion chamber point-dose difference relative to planned dose were investigated. For each patient plan, 5 fractions were delivered. DynaLog files were recorded and centi-MU pulses from dose integrator board for every control point (CP) were counted using a commercial pulse counter. The counter recorded CP MU agrees with DynaLog records, both showing an ~ 0.6MU overshoot of the first segment of every beam. The 3D patient dose was recalculated from the counter records and compared to the planned dose, showing that the overshoot resulted in on average 2.05% of PTV D95 error, and 2.49%, 2.61% and 2.45% of D1cc error for rectum, bladder, and bowel, respectively. The initial plans were then modified by inserting a specially designed MLC segment to the start of every beam. The modified plans were also delivered five times. The dose from the modified delivery was calculated using counter recorded CP MU. The corresponding Dx parameters were all within 0.31% from the original plan. IMRT QA results also show a 2.2% improvement in ion chamber point-dose agreement. The results demonstrate that the proposed plan modification method effectively eliminates the overdosage from the overshooting phenomenon.
منابع مشابه
Step-and-Shoot versus Compensator-based IMRT: Calculation and Comparison of Integral Dose in Non-tumoral and Target Organs in Prostate Cancer
Introduction Intensity-Modulated Radiotherapy (IMRT) is becoming an increasingly routine treatment method. IMRT can be delivered by use of conventional Multileaf Collimators (MLCs) and/or physical compensators. One of the most important factors in selecting an appropriate IMRT technique is integral dose. Integral dose is equal to the mean energy deposited in the total irradiated volume of the p...
متن کاملAssessment and Comparison of Homogeneity and Conformity Indexes in Step-and-Shoot, Compensator-Based Intensity Modulated Radiation Therapy (IMRT) and Three-Dimensional Conformal Radiation Therapy (3D CRT) in Prostate Cancer
Introduction: Intensity modulated radiation therapy (IMRT) and three-dimensional conformal radiation therapy (3D CRT) are two treatment modalities in prostate cancer, which provide acceptable dose distribution in tumor region with sparing the surrounding normal tissues. IMRT is based on inverse planning optimization; in which, intensity of beams is modified by using multileaf c...
متن کاملDosimetric evaluation of nasopharyngeal carcinomas irradiated with different IMRT techniques.
PURPOSE To evaluate the performance of volumetric arc therapy (VMAT), dynamic intensity modulated radiotherapy (IMRT) and step-and-shoot IMRT techniques in nasopharyngeal cancer (NPC) patients. METHODS IMRT plans of 48 NPC patients treated between May 2010 and December 2012, were evaluated. Twenty two patients were planned with VMAT, 18 with dynamic IMRT and 8 with step-and-shoot IMRT. Confor...
متن کاملA study of segment weight optimization with the CMS XiO step‐and‐shoot IMRT technique for prostate cancer
The aim of this study was to compare IMRT optimization in the CMS XiO radiotherapy treatment planning system, with and without segment weight optimization. Twenty-one prostate cancer patients were selected for this study. All patients were initially planned with step-and-shoot IMRT (S-IMRT). A new plan was then created for each patient by applying the segment weight optimization tool (SWO-IMRT)...
متن کاملMonetary policy and exchange rate overshooting in Iran: A Vector Errors Correction (VEC) approach
Assumption of exchange rate overshooting has significant position in international macroeconomic discussion. This phenomenon is one of the abnormal behaviors of exchange rate that happen in short run. Dornbusch (1976) shows that because speed of equilibrium prices is slow relative to asset markets and commodity prices are sticky in the short run, However, over time, commodity prices will rise a...
متن کامل