Intravascular MR imaging-guided balloon angioplasty with an MR imaging guide wire: feasibility study in rabbits.
نویسندگان
چکیده
PURPOSE To develop a technique for intravascular magnetic resonance (MR)-guided balloon angioplasty with use of an MR imaging guide wire. MATERIALS AND METHODS An MR imaging guide wire (0.6-mm loopless antenna) that could be placed within a balloon catheter was manufactured. The guide wire was expected to function as either an MR receiver probe in real-time MR imaging or a guide wire for use with interventional devices. Laparotomy was performed in eight rabbits, and a dilatable stenosis was created at the upper abdominal aorta. Balloon angioplasty, validated at pre- and postoperative MR aortography with renal contrast enhancement was performed by using a 1.5-T MR unit with a fast spoiled gradient-echo pulse sequence, short repetition and echo times, and a rate of three frames per second. RESULTS During MR tracking, the entire length of the MR imaging guide wire was always visible as a band of high signal intensity. In all cases, the MR imaging guide wires were passed through the aortic stenoses dilated by means of balloon inflation. Before balloon angioplasty, flow in the aorta distal to the stenosis was decreased, which caused mild contrast enhancement in each kidney. After balloon angioplasty, distal flow was restored, resulting in substantial renal enhancement. CONCLUSION The MR imaging guide wire is a potential tool for use in endovascular interventional MR imaging.
منابع مشابه
Magnetic resonance-guided cardiac interventions using magnetic resonance-compatible devices: a preclinical study and first-in-man congenital interventions.
BACKGROUND Percutaneous cardiac interventions are currently performed under x-ray guidance. Magnetic resonance imaging (MRI) has been used to guide intravascular interventions in the past, but mainly in animals. Translation of MR-guided interventions into humans has been limited by the lack of MR-compatible and safe equipment, such as MR guide wires with mechanical characteristics similar to st...
متن کاملHigh-resolution real-time spiral MRI for guiding vascular interventions in a rabbit model at 1.5 T.
PURPOSE To study the feasibility of a combined high spatial and temporal resolution real-time spiral MRI sequence for guiding coronary-sized vascular interventions. MATERIALS AND METHODS Eight New Zealand White rabbits (four normal and four with a surgically-created stenosis in the abdominal aorta) were studied. A real-time interactive spiral MRI sequence combining 1.1 x 1.1 mm(2) in-plane re...
متن کاملHigh-resolution intravascular magnetic resonance imaging: monitoring of plaque formation in heritable hyperlipidemic rabbits.
BACKGROUND The individual makeup of atherosclerotic plaque has been identified as a dominant prognostic factor. With the use of an intravascular magnetic resonance (MR) catheter coil, we evaluated the effectiveness of high-resolution MR in the study of the development of atherosclerotic lesions in heritable hyperlipidemic rabbits. METHODS AND RESULTS Sixteen hyperlipidemic rabbits were invest...
متن کاملA Feasibility Study of an Intravascular Imaging Antenna to Image Atherosclerotic Plaques in Swine Using 3.0 T MRI
PURPOSE To investigate the feasibility of an intravascular imaging antenna to image abdominal aorta atherosclerotic plaque in swine using 3.0T magnetic resonance imaging (MRI). METHODS Atherosclerotic model was established in 6 swine. After 8 months, swine underwent an MR examination, which was performed using an intravascular imaging guide-wire, and images of the common iliac artery and the ...
متن کاملIntracoronary MR imaging using a 0.014-inch MR imaging-guidewire: toward MRI-guided coronary interventions.
PURPOSE To validate the feasibility of using a newly designed MR imaging-guidewire (MRIG) to guide angioplasty balloon placement in coronary arteries. MATERIALS AND METHODS A custom gold/sliver/Nitinol/MP35N-based, 0.014-inch MRIG was manufactured. To test its mechanical performance we used the new MRIG to catheterize the left coronary arteries of three dogs under x-ray fluoroscopy. To furthe...