The impact of spectral filtration on image quality in micro‐CT system

نویسندگان

  • Liqiang Ren
  • Muhammad U Ghani
  • Di Wu
  • Bin Zheng
  • Yong Chen
  • Kai Yang
  • Xizeng Wu
  • Hong Liu
چکیده

This paper aims to evaluate the impact of spectral filtration on image quality in a microcomputed tomography (micro-CT) system. A mouse phantom comprising 11 rods for modeling lung, muscle, adipose, and bones was scanned with 17 s and 2 min, respectively. The current (μA) for each scan was adjusted to achieve identical entrance exposure to the phantom, providing a baseline for image quality evaluation. For each region of interest (ROI) within specific composition, CT number variations, noise levels, and contrast-to-noise ratios (CNRs) were evaluated from the reconstructed images. CT number variations and CNRs for bone with high density, muscle, and adipose were compared with theoretical predictions. The results show that the impact of spectral filtration on image quality indicators, such as CNR in a micro-CT system, is significantly associated with tissue characteristics. The findings may provide useful references for optimizing the scanning parameters of general micro-CT systems in future imaging applications. PACS numbers: 87.57.C-, 87.57.Q-, 87.64.kd.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment the effect of CT localizer parameter on radiation dose and image quality of GE CT spiral images, A phantom study

Introduction: Using computed tomography (CT) is increasing significantly in Iran and worldwide. Reduction of CT's radiation dose is a critical concern. The aim of this study was evaluating the effect of localizer kVp on spiral CT images when the tube current modulation (TCM) is activated. Materials and Methods: A special wedding cake phantom with circular in...

متن کامل

Investigating the effects of different kernels used for CT image reconstruction on dose distributions in treatment planning of kidney cancer radiotherapy

Introduction: The quality of CT images used for treatment planning of cancer patients is an important issue in accurate outlining of the tumor volume and organs at risk. Different kernels in CT scanner systems are available for improving the image quality. Applying these kernels  on   CT  images  will  change   the   CT  numbers  and  electron   density   of   tissues, conse...

متن کامل

An assessment of the Doses Received by Children from CT Examinations Along with the QC Parameters from a Conventional CT System

Introduction: In 2000, the UNSCEAR reported that CT constitutes 5% of all the medical x-ray  examinations and it contributes 34% of the resultant collective dose worldwide. Children are more  sensitive to the ionizing radiations than adults. So, routine quality control tests are expected to be  carried out periodically on the CT scanners. The aim of this research was to estimate the effective  ...

متن کامل

Impact of Novel Incorporation of CT-based Segment Mapping into a Conjugated Gradient Algorithm on Bone SPECT Imaging: Fundamental Characteristics of a Context-specific Reconstruction Method

Objective(s): The latest single-photon emission computed tomography (SPECT)/computed tomography (CT) reconstruction system, referred to as xSPECT Bone™, is a context-specific reconstruction system utilizing tissue segmentation information from CT data, which is called a zone map. The aim of this study was to evaluate theeffects of zone-map enhancement incorporated into the ordered-subset conjug...

متن کامل

Application of Different methods for Reducing Radiation Dose to Breast during MDCT

The increased use of computed tomography (CT) and its high radiation dose have led to great concerns about its potential for radiation induced cancer risks. Breast is a radiosensitive tissue based on tissue weighting factors assigned by the International Commission on Radiological Protection (ICRP). Moreover, the dose is maximal on the surface of the patient. Therefore, strategies should be tak...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 17  شماره 

صفحات  -

تاریخ انتشار 2016