Convex gradient optimization for increased spatiotemporal resolution and improved accuracy in phase contrast MRI

نویسندگان

  • Matthew J Middione
  • Holden H Wu
  • Daniel B Ennis
چکیده

PURPOSE To evaluate convex gradient optimization (CVX) for increased spatiotemporal resolution and improved accuracy for phase-contrast MRI (PC-MRI). METHODS A conventional flow-compensated and flow-encoded (FCFE) PC-MRI sequence was compared with a CVX PC-MRI sequence using numerical simulations, flow phantom experiments, and in vivo experiments. Flow measurements within the ascending aorta, main pulmonary artery, and right/left pulmonary arteries of normal volunteers (N = 10) were acquired at 3T and analyzed using a conventional FCFE sequence and a CVX sequence with either higher spatial resolution or higher temporal resolution. All sequences mitigated chemical shift-induced phase errors and used equivalent breath-hold durations. RESULTS Chemical shift-optimized PC-MRI has increased sequence efficiency when using CVX, which can provide either higher spatial or higher temporal resolution compared with conventional FCFE PC-MRI. Numerical simulations, flow phantom experiments, and in vivo experiments indicate that CVX measurements of total flow and peak velocity are increased and more accurate when compared with FCFE. CONCLUSION CVX PC-MRI increases sequence efficiency while reducing chemical shift-induced phase errors. This can be used to provide either higher spatial or higher temporal resolution than conventional chemical shift-mitigated PC-MRI methods to provide more accurate measurements of blood flow and peak velocity.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Rapid phase contrast MRI with minimum time gradient waveform design using convex optimization

Background Phase contrast MRI (PC-MRI) sequences conventionally use triangular and trapezoidal gradient waveforms to construct flow compensated and flow encoded (FCFE) [1] or Bipolar [2] velocity encoding gradients. This approach does not make optimal use of the available gradient hardware for all time points, but convex optimization (CVX) can be used to design minimum-duration, arbitrarily sha...

متن کامل

Particle Swarm Optimization for Hydraulic Analysis of Water Distribution Systems

The analysis of flow in water-distribution networks with several pumps by the Content Model may be turned into a non-convex optimization uncertain problem with multiple solutions. Newton-based methods such as GGA are not able to capture a global optimum in these situations. On the other hand, evolutionary methods designed to use the population of individuals may find a global solution even for ...

متن کامل

Spatiotemporal Estimation of PM2.5 Concentration Using Remotely Sensed Data, Machine Learning, and Optimization Algorithms

PM 2.5 (particles <2.5 μm in aerodynamic diameter) can be measured by ground station data in urban areas, but the number of these stations and their geographical coverage is limited. Therefore, these data are not adequate for calculating concentrations of Pm2.5 over a large urban area. This study aims to use Aerosol Optical Depth (AOD) satellite images and meteorological data from 2014 to 2017 ...

متن کامل

Model-based reconstruction for real-time phase-contrast flow MRI: Improved spatiotemporal accuracy.

PURPOSE To develop a model-based reconstruction technique for real-time phase-contrast flow MRI with improved spatiotemporal accuracy in comparison to methods using phase differences of two separately reconstructed images with differential flow encodings. METHODS The proposed method jointly computes a common image, a phase-contrast map, and a set of coil sensitivities from every pair of flow-...

متن کامل

Motion Compensated Dynamic Imaging without Explicit Motion Estimation

Purpose – The main focus of this abstract is to recover dynamic MRI data from highly under-sampled measurements. Compressed sensing schemes that exploit sparsity in Fourier and gradient domains have enjoyed a lot of success in breath-held cardiac MRI. However, these schemes often result in un-acceptable spatiotemporal blurring and residual alias artifacts, when applied to free breathing cardiac...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2014