Ion chamber measurements of transverse gamma knife beam profiles

نویسنده

  • Morris I. Bank
چکیده

A microchamber, PTW Pinpoint 31006, was used to measure transverse beam profiles for an Elekta Gamma Knife, Model B, and compared with profiles measured with film dosimetry. The microchamber sensitive volume has a diameter of 2 mm, which is smaller than the gamma knife beams, and a length of 5 mm. The chamber was mounted in a custom cassette in a spherical plastic phantom, supplied by Elekta, and oriented in a sagittal plane with the 2-mm dimension at right angles to the transverse plane. The phantom was manually moved across the beam, using the gamma knife x-coordinate trunnions, to measure the profiles. Profiles were also measured with V-film placed in a cassette mounted in the spherical plastic phantom. The films were scanned with a Scanditronix film scanner and converted to dose with a density to dose calibration curve. The results were superimposed for comparison. The beam width at the 50% intensity was measured from the film profiles to give the dimensions of the beams in the orthagonal planes. The ion chamber measurements are compared with the film results for the transverse, x profiles. Good agreement between the film and ion chamber transverse profiles is observed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Superimposition of beams to vary shot size in gamma knife stereotactic radiosurgery

The Leksell Gamma Knife [Elekta Corp] uses helmets as collimators to produce four standard beam sizes. The nominal beam diameters are 18, 14, 8, and 4 mm. During computer treatment planning for gamma knife stereotactic radiosurgery, the size of the treated volume may differ from the standard beam sizes. To maintain conformality of the isodose curves to the treated volume, beam sizes may be supe...

متن کامل

Study the effect of source to diaphragm distance (SDD) on beam penumbra width of Gamma Knife machine model 4C using Monte Carlo simulation

Introduction: Stereotactic radiosurgery (SRS) is one of the conformal radiotherapy methods and Gamma Knife is a non-invasive SRS technique to treat intracranial lesions with 201 Co- 60 sources without need to surgery and bleeding. Beam penumbra phenomenon leads to irradiation outside the field and normal tissues. Therefore, considering it is momentous to attain the acceptable t...

متن کامل

Penumbra Measurements and Comparison of In-House and Standard Circular Cones by the Gafchoromic Film, Pinpoint Ion Chamber, and MCNPX Monte Carlo Simulation

Introduction: Penumbra is an important property of the radiation beam to obtain a suitable margin surrounding the target volume. Therefore, the precise penumbra width determination in stereotactic radiotherapy is necessary for treatment planning. This study aimed to compare the obtained results of penumbra width by in-house and standard circul...

متن کامل

Evaluation of the Effects of Inhomogeneities on Dose Profiles Using Polymer Gel Dosimeter and Monte Carlo Simulation in Gamma Knife

Introduction Polymer gel dosimeters offer a practical solution to 3D dose verification for conventional radiotherapy as well as intensity-modulated and stereotactic radiotherapy. In this study, EGSnrc calculated and PAGAT polymer gel dosimeter measured dose profiles from single shot irradiation with 18 mm collimator of Gamma Knife in homogeneous and inhomogeneous phantoms were compared with eac...

متن کامل

Monte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system.

The accuracy of single-beam dose profiles used in the algorithm of the Gamma Knife treatment planning system (Leksell GammaPlan) is verified. EGS4 Monte Carlo calculation was employed to calculate the dose distributions of single-beams in a spherical water phantom with diameter 160 mm. The beams were directed to the center of the phantom. Collimators of 4, 8, 14, and 18 mm sizes were studied. T...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 3  شماره 

صفحات  -

تاریخ انتشار 2002