Comparison of sound quality and clarity with asymmetrical peak clipping and output limiting compression.

نویسندگان

  • D B Hawkins
  • S V Naidoo
چکیده

This study compared the preferences of 12 subjects with mild-to-moderate hearing losses for a linear circuit when the type of output limitation could be changed between asymmetrical peak clipping and output limiting compression. Through a paired-comparison paradigm, subjects rated the sound quality and clarity of speech in quiet, speech in noise, and music when the output levels were well below saturation and when slight saturation and high saturation occurred. A significant preference was found for the output limiting compression for each of the three stimuli for both sound quality and clarity. The preference became stronger with increases in saturation. The results suggest that if linear hearing aids are being used, the output should be limited with compression rather than peak clipping.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monaural/binaural preferences: effect of hearing aid circuit on speech intelligibility and sound quality.

This investigation compared monaural and binaural hearing aid preferences of 15 adults with mild-to-moderately-severe bilaterally symmetrical sensorineural hearing losses. Subjects listened to connected discourse in quiet and background noise at 70 and 80 dB SPL with K-Amp, linear Class D, linear output limiting compression (OLC), Manhattan II, and linear asymmetrical peak clipping circuits (AP...

متن کامل

Ranking hearing aid input-output functions for understanding low-, conversational-, and high-level speech in multitalker babble.

PURPOSE To determine the rankings of 6 input-output functions for understanding low-level, conversational, and high-level speech in multitalker babble without manipulating volume control for listeners with normal hearing, flat sensorineural hearing loss, and mildly sloping sensorineural hearing loss. METHOD Peak clipping, compression limiting, and 4 wide dynamic range compression (WDRC) input...

متن کامل

An evidence-based systematic review of amplitude compression in hearing aids for school-age children with hearing loss.

PURPOSE Two clinical questions were developed: one addressing the comparison of linear amplification with compression limiting to linear amplification with peak clipping, and the second comparing wide dynamic range compression with linear amplification for outcomes of audibility, speech recognition, speech and language, and self- or parent report in children with hearing loss. METHOD Twenty-s...

متن کامل

Hearing-aid-processed tone pips: electroacoustic and ABR characteristics.

The auditory brainstem response (ABR) recorded while wearing a hearing aid may supply supplemental information about the benefit and appropriateness of the hearing aid for certain infants. The purposes of this study were (1) to determine the effects of different output limiting circuits on the acoustics of tone-pip stimuli used for ABR recordings and (2) assess how changes in hearing-aid-proces...

متن کامل

Effect of preferred volume setting on speech audibility in different hearing aid circuits.

This study compared preferred volume setting for linear peak clipping, compression limiting, and wide dynamic range compression (WDRC) amplification and quantified speech audibility at the preferred volume setting for each amplification type. Ten listeners with mild-to-moderate hearing loss were fitted monaurally with a behind-the-ear hearing aid programmed sequentially with WDRC, compression l...