Ultra fast, accurate PET image reconstruction for the Siemens hybrid MR/BrainPET scanner using raw LOR data
نویسندگان
چکیده
Forschungszentrum Jülich GmbH, Jülich, Germany Fast PET image reconstruction algorithms usually use a Line-of-Response (LOR) preprocessing step where the detected raw LOR data are interpolated either to evenly spaced sinogram projection bins or alternatively to a generic projection space as for example proposed by the PET Reconstruction Software Toolkit (PRESTO) [1]. In this way, speed-optimised, versatile geometrical projectors can be implemented for iterative image reconstruction independent of the underlying scanner geometry. However, all strategies of projection data interpolation unavoidably lead to a loss of original information and result in some degradation of image quality. Here, direct LOR reconstructions overcome this evident drawback at cost of a massively enhanced computational burden. Therefore, computational optimisation techniques are essential to make such demanding approaches attractive and economical for widespread usage in the clinical environment. In this paper, we demonstrate for the Siemens Hybrid MR/BrainPET with 240 million physical LORs that a very fast quantitative direct LOR reconstruction can be realized using a modified version of PRESTO. Now, PRESTO is also capable to directly use sets of symmetric physical LORs instead of interpolating LORs to a generic projection space. Exploiting basic scanner symmetries together with the technique of Single Instruction Multipe Data (SIMD) and Simultaneous Multi-Threading (SMT) results in an overall calculation time of 2-3 minutes per frame on a single multi-core machine, i.e. neither requiring a cluster of mutliple machines nor Graphics Processing Units (GPUs).
منابع مشابه
High-resolution, quantitative 3D PET image reconstruction for the Siemens hybrid 3T MR/BrainPET scanner using the PET reconstruction software toolkit (PRESTO)
Institute of Neuroscience and Medicine-4, Forschungszentrum Juelich, Germany The Siemens 3T MR-BrainPET scanner allows us to simultaneously acquire high-resolution MR and PET images thus giving a strong asset for studies of the human brain. Meanwhile, the system is routinely used for MR-PET studies with a variety of radiotracers, e.g. F-FDG, F-FET, C-Raclopride, C-Flumazenil, O-Water. Based on ...
متن کاملCombined MR-assisted motion and partial volume effects corrections – impact on PET data quantification
Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Charlestown, MA, USA Our goal in this study was to characterize the combined effect of MR-assisted motion correction (MC) and partial volume effects correction (PVEC) on the estimation of [C]NNC112 binding potential (BP) in healthy volunteers. 29 subjects were...
متن کاملImage reconstruction of mMR PET data using the open source software STIR
Centre for Medical Image Computing, University College London, London, UK Simultaneous PET and MR acquisitions have now become possible with the new hybrid Biograph Molecular MR (mMR) scanner from Siemens. The purpose of this work is to create a platform for mMR 3D and 4D PET image reconstruction which would be freely accessible to the community as well as fully adjustable in order to obtain op...
متن کاملMRI-assisted PET motion correction for neurologic studies in an integrated MR-PET scanner.
UNLABELLED Head motion is difficult to avoid in long PET studies, degrading the image quality and offsetting the benefit of using a high-resolution scanner. As a potential solution in an integrated MR-PET scanner, the simultaneously acquired MRI data can be used for motion tracking. In this work, a novel algorithm for data processing and rigid-body motion correction (MC) for the MRI-compatible ...
متن کاملPET motion correction using PRESTO with ITK motion estimation
Institute of Biophysics and Biomedical Engineering, Science Faculty of University of Lisbon, Portugal The Siemens BrainPET scanner is a hybrid MRI/PET system. PET images are prone to motion artefacts which degrade the image quality. Therefore, motion correction is essential. The library PRESTO converts motion-corrected LORs into highly accurate generic projection data [1], providing high-resolu...
متن کامل