Spectrophotometric determination of iron and boron in soil extracts using a multi-syringe flow injection system.

نویسندگان

  • D M C Gomes
  • M A Segundo
  • J L F C Lima
  • A O S S Rangel
چکیده

In the last decade, significant advances in flow analysis have been reported, namely the extensive use of computer-controlled devices to enhance the autonomy and performance of analysers. In the present work, computer-controlled multi-syringe flow injection systems are proposed to perform the spectrophotometric determination of available iron and boron in soil extracts. The methodologies were based on the formation of ferroin complex (determination of iron) and azomethine-H reaction (determination of boron). Both determinations were performed in manifolds with similar configurations by changing the reagents present in the different syringes. In the determination of iron, elimination of Schlieren effect in the detection system was achieved through the binary sampling approach, where a three-way valve was actuated to intercalate small slugs of sample and reagent, promoting better mixing conditions for solutions with different values of refractive index. In the determination of boron, in-line sample blank measurement was attained by omitting the introduction of reagent through software control, without manifold reconfiguration. Linear calibration curves were established between 0.50 and 10.0mgFel(-1) and between 0.20 and 4.0mgBl(-1). No systematic difference was found when soil extracts were analysed by the proposed methodologies and compared to the respective reference procedures.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A multi-syringe flow injection system for the spectrophotometric determination of trace levels of iron in waters using a liquid waveguide capillary cell and different chelating resins and reaction chemistries

A method integrating a long waveguide capillary cell with a preconcentration resin in a multi-syringe flow injection analysis (MSFIA) system for iron determination in waters was developed. The determination of iron is based on a colorimetric reaction and two reagents were tested, ferrozine and ammonium thiocyanate. A liquid waveguide capillary cell (1.0 m pathlength, 550 μm i.d. and 250 μL inte...

متن کامل

Potentiometric multi-syringe flow of exchangeable potassium in soils with in-line extraction

A multi-syringe flow injection system for the potentiometric determination of exchangeable potassium in soil samples is proposed. Firstly, a manifold was devised to allow determination in soil extracts prepared off-line. It was possible to analyze samples prepared in extractants with different composition (Mehlich or Morgan) without physical or chemical modification of the manifold. A linear dy...

متن کامل

Flow Injection Spectrophotometric Determination of Sulphide by Catalytic Reduction of Toluidine Blue

Sulphide was determined spectrophotometrically at 628 nm and pH 6.7 based on its catalytic reduction of toluidine blue in the presence of Se(IV) using a flow injection system. After optimization of the measuring conditions the calibration curve was found linear up to 25 mg/mL and the detection limit was 0.3 mg/mL (30 mL injections). The ...

متن کامل

Spectrophotometric determination of zinc and copper in a multi-syringe flow injection analysis system using a liquid waveguide capillary cell: application to natural waters.

This work exploits a multi-syringe injection analysis (MSFIA) system coupled with a long liquid waveguide capillary cell for the spectrophotometric determination of zinc and copper in waters. A liquid waveguide capillary cell (1.0m pathlength, 550 μm i.d. and 250 μL internal volume) was used to enhance the sensitivity of the detection. The determination for both ions is based on a colorimetric ...

متن کامل

Ultra-trace Determination of Palladium(II) by Spectrophotometric Flow Injection Analysis

Nowadays, palladium compounds due to their especial characteristics used in different industries, mainly in electrical power sources and chemical productions as a catalyst. Therefore, Pd concentration is increasing in our around that can be contaminant our environment due its toxic effects. In this work, a simple, selective and rapid flow injection method of analysis (FIA) has been developed fo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Talanta

دوره 66 3  شماره 

صفحات  -

تاریخ انتشار 2005