Poor agreement between a portable armband and indirect calorimetry in the assessment of resting energy expenditure.

نویسندگان

  • S Bertoli
  • A Posata
  • A Battezzati
  • A Spadafranca
  • G Testolin
  • G Bedogni
چکیده

BACKGROUND & AIMS To evaluate the agreement between resting energy expenditure (REE) estimated by a portable armband and measured by indirect calorimetry. METHODS One-hundred and twenty-seven women and 42 men with a mean (SD) age of 44 (12) years and a body mass index of 30.2 (5.4) kg/m(2) were studied. REE was estimated using the Sense Wear Pro 2 Armband (SWA), measured using the Sensor Medics 29 metabolic cart (V(max)), and estimated using Schofield's equation. The limits of agreement (LOA) and the concordance correlation coefficient (CCC) were used to evaluate the interchangeability of the methods. RESULTS The LOA between REE(SWA) and REE(Vmax) were wide in both women (-269 to 378 kcal/day) and men (-330 to 545 kcal/day) and CCC was low (0.579 in females and 0.583 in males, p<0.0001 for both). REE(Schofield) agreed with REE(Vmax) to a similar degree (CCC=0.563 in females and 0.500 in males, p<0.0001 for both). CONCLUSIONS SWA and indirect calorimetry are not interchangeable methods for the assessment of REE in normal-weight and obese subjects.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a portable device to measure daily energy expenditure in free-living adults.

BACKGROUND Increasing daily energy expenditure (EE) plays an important role in the prevention or treatment of several lifestyle-related diseases; however, its measurement remains problematic. OBJECTIVE The objective was to evaluate a portable armband device for measuring daily and physical activity EE compared with doubly labeled water (DLW) in free-living individuals. DESIGN Daily EE and p...

متن کامل

Accuracy of a portable multisensor body monitor for predicting resting energy expenditure in older people: a comparison with indirect calorimetry.

BACKGROUND Accurate and comfortable methods are needed to determine resting energy expenditure (REE) in older people who are characterized by a lowered metabolic rate. The portable SenseWear® armband (SWA) body monitor, worn on the right upper arm, can easily be used by this age group in an ambulatory manner. OBJECTIVE The purpose of this study was to evaluate the reliability and accuracy of ...

متن کامل

[Analysis of energy expenditure in adults with cystic fibrosis: comparison of indirect calorimetry and prediction equations].

OBJECTIVE Undernutrition, which implies an imbalance between energy intake and energy requirements, is common in patients with cystic fibrosis. The aim of this study was to compare resting energy expenditure determined by indirect calorimetry with that obtained with commonly used predictive equations in adults with cystic fibrosis and to assess the influence of clinical variables on the values ...

متن کامل

Energy Expenditure Evaluation in Humans and Non-Human Primates by SenseWear Armband. Validation of Energy Expenditure Evaluation by SenseWear Armband by Direct Comparison with Indirect Calorimetry

INTRODUCTION The purpose of this study was to compare and validate the use of SenseWear Armband (SWA) placed on the arm (SWA ARM) and on the back (SWA BACK) in healthy humans during resting and a cycle-ergometer exercise and to evaluate the SWA to estimate Resting Energy Expenditure (REE) and Total Energy Expenditure (TEE) in healthy baboons. METHODS We studied 26 (15F/11M) human subjects wea...

متن کامل

Validation of a predictive method for an accurate assessment of resting energy expenditure in medical mechanically ventilated patients.

OBJECTIVE Use comparison with indirect calorimetry to confirm the ability of our previously described equation to predict resting energy expenditure in mechanically ventilated patients. DESIGN Prospective, validation study. SETTING Eighteen-bed, medical intensive care unit at a teaching hospital. PATIENTS All adult patients intubated >24 hrs were assessed for eligibility. Exclusion criter...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical nutrition

دوره 27 2  شماره 

صفحات  -

تاریخ انتشار 2008