Identification of microemboli during haemodialysis using Doppler ultrasound.

نویسندگان

  • F Rollé
  • J Pengloan
  • M Abazza
  • J M Halimi
  • M Laskar
  • L Pourcelot
  • F Tranquart
چکیده

BACKGROUND Doppler ultrasound methods were used during haemodialysis sessions for the detection of microemboli and determination of their origin. METHODS A 2-MHz ultrasound probe (Multidop X(4) DWL((TM))) was used to assess the number of microembolic signals (MES) in the subclavian vein downstream from the arteriovenous fistula before the dialysis session and over two periods of 15 min at the beginning and end of haemodialysis sessions in 25 patients without previous cardiovascular disease. A similar probe was used during in vitro studies to detect MES at different sites in the dialysis machine (before and downstream from the blood pump, and before and downstream from the air trap). RESULTS No MES were detected during in vivo studies before haemodialysis sessions. MES were registered in all patients (100%) at the beginning and end of the haemodialysis procedure at an average of 12.7+/-9 and 16. 7+/-11.5 signals/min respectively. The average intensity of MES was 19.2+/-5.0 dB and 19.4+/-3.9 dB respectively. No MES were detected on the arterial line during in vitro studies. In contrast, 19+/-6 MES/min were detected after the blood pump, 13+/-4.2 before the air trap, and 16.5+/-5.5 thereafter. CONCLUSIONS In all patients, MES were recorded during haemodialysis sessions in the drainage vein from arteriovenous fistulae. The results of in vitro studies indicate that MES are formed by the blood pump of the haemodialysis machine. The intensity of the MES suggests that they correspond to synthetic particles or microbubbles, which are not detected by the air trap. The final destination of these microbubbles will be assessed in further studies.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Detection of microemboli in the subclavian vein of patients undergoing haemodialysis and haemodiafiltration using pulsed Doppler ultrasound.

BACKGROUND The pathophysiology leading to pulmonary side effects during haemodialysis and haemodiafiltration is not yet fully understood. Chronic microembolization, which can be demonstrated by pulsed Doppler ultrasound, may be one cause. METHODS The study cohort consisted of 24 long-term dialysis patients undergoing haemodialysis (n=21) and online-haemodiafiltration (n=3), respectively. The ...

متن کامل

Reduction of circulating microemboli in the subclavian vein of patients undergoing haemodialysis using pre-filled instead of dry dialysers.

BACKGROUND Chronic microembolization that can be demonstrated by pulsed Doppler ultrasound may give rise to pulmonary side-effects during haemodialysis by direct vessel obstruction, increased complement activation or platelet aggregation. The objective of the present investigation was to study whether the use of pre-filled instead of dry dialysers would help to minimize the number of microembol...

متن کامل

Detection of Doppler Microembolic Signals Using High Order Statistics

Robust detection of the smallest circulating cerebral microemboli is an efficient way of preventing strokes, which is second cause of mortality worldwide. Transcranial Doppler ultrasound is widely considered the most convenient system for the detection of microemboli. The most common standard detection is achieved through the Doppler energy signal and depends on an empirically set constant thre...

متن کامل

Pii: S0301-5629(01)00486-0

Difficulties in location of transcranial ultrasound (US) windows and blood flow in cerebral vessels, and unambiguous detection of microemboli, have limited expansion of transcranial Doppler US. We developed a new transcranial Doppler modality, power M-mode Doppler (PMD), for addressing these issues. A 2-MHz digital Doppler (Spencer Technologies TCD100M) having 33 sample gates placed with 2-mm s...

متن کامل

Review of transcranial Doppler ultrasound to detect microemboli during orthopedic surgery.

Transcranial Doppler ultrasonography has been used to detect microemboli in the middle cerebral artery during orthopedic surgery. We conducted a comprehensive systematic literature review of transcranial Doppler ultrasonography in orthopedic surgery to evaluate its status in this setting. Fourteen studies were selected for qualitative analysis. The highest number of patients studied was 45; emb...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Nephrology, dialysis, transplantation : official publication of the European Dialysis and Transplant Association - European Renal Association

دوره 15 9  شماره 

صفحات  -

تاریخ انتشار 2000