Clinical evaluation of a new saturation/hematocrit monitor.

نویسندگان

  • M F Miller
  • J Luckenbach
  • C Chen
چکیده

The purpose of this study was to evaluate the utility of the Cobe SAT/HCT Monitor, an in-line device which continuously displays hematocrit and venous oxygen saturation. A fiber optic cable was used to connect the monitor to a "window" on the blood inlet of a venous reservoir (Cobe CML). Sixty venous blood samples were drawn from 20 patients undergoing cardiopulmonary bypass. Statistical correlations were performed comparing monitor readings with controls measured by an IL 482 co-oximeter and a TRIAC centrifuge. Results showed a high degree of correlation between the monitor and control values for both venous saturation (r = .96) and for hematocrit (r = .94). Although this device has some minor deficiencies, it is a useful tool which enables the perfusionist to monitor two important components of oxygen consumption.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Clinical evaluation of the CDI-100 in-line hematocrit/saturation monitor.

This study was undertaken to evaluate the accuracy, reliability, consistency and biases of the CDI-100 saturation monitor when compared with a blood gas analyzer. The advantage of continuous in-line monitoring is that the perfusionist has continuous updates as to the patient's changing physiologic state. During this study, if the sample readout of the CDI-100 was off by greater than 10% from th...

متن کامل

In vitro evaluation of continuous mixed venous oxygen saturation and hematocrit monitors.

Four venous hemoglobin oxygen saturation monitors, two of which provide hemoglobin monitoring, were evaluated and compared to a performance standard (control) during cardiopulmonary bypass (CPB). The CDI 100 Hematocrit/Saturation Monitor, CDI 400, Cobe Hematocrit/Saturation Monitor, and the Baxter Bentley OxySat values were compared to the Radiometer Copenhagen OSM3 Hemoximeter and the Damon In...

متن کامل

Clinical comparisons of continuous venous oxygen saturation and hematocrit monitors in pediatric surgery.

Continuous venous oxygen saturation and hematocrit values are important parameters in assessing patient status while on cardiopulmonary bypass. Two devices used to measure continuous venous oxygen saturation while on cardiopulmonary bypass were compared to a control. The Bentley Oxysat meter and the Medtronic MX2 Oxygen Saturation and Hematocrit System were compared to the ABL500 blood gas mon...

متن کامل

Clinical evaluation of a new in-line continuous blood gas monitor.

Two methodologies for obtaining accurate blood gas and electrolyte values during cardiopulmonary bypass (CPB) are traditional laboratory analyzers, which use an electrochemical technology, and continuous in-line monitoring systems, which use a fluorometric and/or spectrophotometric technology. The purpose of the present study was to evaluate the accuracy of a new continuous in-line monitor, the...

متن کامل

Continuous venous oximetry: a comparative study between the CDI 100 and the Bentley OxySat II.

Two in-line oxygen saturation monitors, the CDI 100 and OxySat II, were evaluated in the clinical setting. Eighty-seven venous blood samples were drawn during 20 elective cardiopulmonary bypass procedures. Monitor readings were compared to OSM III co-oximeter values. The results revealed that saturation (%) determination was biased, -3.16 +/- 2.21 SD for the CDI 100 and -0.34 +/- 2.17 SD for th...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The journal of extra-corporeal technology

دوره 24 2  شماره 

صفحات  -

تاریخ انتشار 1992