Exchangeable Femoral Neck (Dual-Modular) THA Prostheses Have Poorer Survivorship Than Other Designs: A Nationwide Cohort of 324,108 Patients

نویسندگان

  • Sandrine Colas
  • Assia Allalou
  • Antoine Poichotte
  • Philippe Piriou
  • Rosemary Dray-Spira
  • Mahmoud Zureik
چکیده

BACKGROUND Exchangeable neck stems, defined as those with a dual taper (that is, a modular junction between the femoral head and the femoral neck and an additional junction between the neck and the stem body), were introduced in THA to improve restoration of joint biomechanics (restoring anteversion, offset, and limb length) and reduce the risk of dislocation. However exchangeable necks have been reported to result in adverse effects such as stem fractures and acute local tissue reaction. Whether they result in a net improvement to or impairment of reconstructive survivorship remains controversial. QUESTIONS/PURPOSES (1) To compare the prosthetic survivorship and all-cause revision risk of exchangeable femoral neck THAs versus fixed neck THAs, taking known prosthetic revision risk factors into account; and (2) to compare the cause-specific revision risk of exchangeable femoral neck THAs versus fixed neck THAs, adjusting for known prosthetic risk factors. METHODS Using French national health-insurance databases, we identified all French patients older than 40 years who underwent primary THA from 2009 through 2012. To ensure accuracy of the data, we considered only beneficiaries of the general insurance scheme (approximately 77% of the population). Characteristics of the prosthesis and the patients receiving an exchangeable femoral neck THA were compared with those receiving a fixed femoral neck THA (defined as femoral stem with only the head being exchangeable). Revision was the event of interest. Followup started on the date the THA was performed, until the patient experienced revision, died, was lost to followup, or until the followup period ended (December 31, 2014), whichever came first. Competing risk THA survivorship was calculated and compared (purpose 1), as were cause-specific Cox regression models (purpose 2). The study cohort included 324,108 individuals with a mean age of 77 years. A total of 24% underwent THA for acute trauma, and 3% of the group received an exchangeable neck THA. During the median 45-month followup (mean, 42 months; minimum, 1 day; maximum, 6 years), 11,968 individuals underwent prosthetic revision. RESULTS The cumulative revision incidence was 6.5% (95% CI, 5.8%-7.3%) for exchangeable neck THAs versus 4.7% (95% CI, 4.6%-4.8%) for fixed neck THAs (p < 0.001). After controlling for potential confounding variables including age, sex, comorbidities, indication for THA, cementation, bearing surface, and the characteristics of the center where the implantation was performed, we found that the exchangeable femoral neck THA was associated with an increased hazard ratio (HR) of revision of 1.26 (95% CI, 1.14-1.38; p < 0.001) compared with the fixed neck THA. When dealing with cause-specific revision, exchangeable neck THAs had a higher incidence of revision for implant failure or periprosthetic fracture, and for mechanical complications; adjusted HRs were, respectively, 1.68 (95% CI, 1.24-2.27; p < 0.001) and 1.27 (95% CI, 1.13-1.43; p < 0.001), for exchangeable neck THAs compared with fixed ones. CONCLUSIONS Exchangeable neck THAs had poorer survivorship independent of other prosthetic revision risk factors. Accordingly, expected anatomic and functional benefits should be carefully assessed before choosing this design. LEVEL OF EVIDENCE Level III, therapeutic study.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cementless Hip Arthroplasty in Southern Iran, Midterm Outcome and Comparison of Two Designs

Background: Cementless hip prosthesis was designed to provide biologic fixation, without the use of cement. The second generation components have shown more reliable bone ingrowths and survival rates. We are reporting a midterm result of two designs of cementless prosthesis in a unique culture with different social habits and expectations.Methods: 52 primary cementless total hip arthroplasty in...

متن کامل

Early mortality and morbidity after total hip arthroplasty in patients with femoral neck fracture

Background and purpose - Early postoperative mortality is relatively high after total hip arthroplasty (THA) that has been performed due to femoral neck fracture. However, this has rarely been investigated after adjustment for medical comorbidity and comparison with the mortality in an age-matched population. We therefore assessed early mortality in hip fracture patients treated with a THA, in ...

متن کامل

Will the Clinical Outcome of Cemented Modular Neck Total Hip Prostheses be Comparable to Conventional Devices?

INTRODUCTION: Modular neck total hip arthroplasty (THA) devices afford the surgeon increased intra-operative flexibility in terms of the degree of head version and offset, as well as allowing more controlled stem positioning. Strains on the outer cortical bone surface have been shown to be strongly dependent on the orientation of a THA. These strains can be a potential cause for mechanical fail...

متن کامل

Novel concept of a modular hip implant could contribute to less implant failure in THA: a hypothesis

Background The modularity in total hip arthroplasty (THA) allows orthopaedic surgeons for an exact reconstruction of hip biomechanical parameters especially in revision and tumor arthroplasty. Modular structured femoral stems using taper junctions showed increased implant breakage in the recent past. Presentation of the hypothesis We hypothesize that a novel modular stem-neck-interface leads ...

متن کامل

Poor Survivorship and Frequent Complications at a Median of 10 Years After Metal-on-Metal Hip Resurfacing Revision

BACKGROUND High short-term failure rates have been reported for several metal-on-metal hip resurfacing (MoMHR) designs. Early observations suggested that MoMHRs revised to total hip arthroplasties (THAs) for pseudotumor had more major complications and inferior patient-reported outcomes compared with other revision indications. However, little is known about implant survivorship and patient-rep...