Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification

نویسندگان

  • Sathiyan Saminathan
  • Ravikumar Manickan
  • Varatharaj Chandraraj
  • Sanjay. S. Supe
چکیده

Intensity-modulated radiotherapy treatment demands stringent quality assurance and accurate dose determination for delivery of highly conformal dose to the patients. Generally 3D dose distributions obtained from a treatment planning system have to be verified by dosimetric methods. Mainly, a comparison of two-dimensional calculated and measured data in several coplanar planes is performed. In principle, there are many possibilities to measure two-dimensional dose distributions such as films, flat-panel electronic portal imaging devices (EPID), ion chambers and ionization chamber arrays, and radiographic and radiochromic films. The flat-panel EPIDs show a good resolution and offer a possibility for real-time measurements: however to convert the signal into dose, a separate commercial algorithm is required. The 2D ion chamber array system offers the real-time measurements. In this study, dosimetric characteristics of 2D ion chamber array matrix were analyzed for verification of radiotherapy treatments. The dose linearity and dose rate effect of the I'matriXX device was studied using 6 MV, 18 MV photons and 12 MeV electrons. The output factor was estimated using I'matriXX device and compared with ion chamber measurements. The ion chamber array system was found to be linear in the dose range of 2-500 cGy and the response of the detector was found to be independent of dose rate between 100 MU/min to 600 MU/min. The estimated relative output factor with I'matriXX was found to match very well with the ion chamber measurements. To check the final dose delivered during IMRT planning, dose distribution patterns such as field-in-field, pyramidal, and chair tests were generated with the treatment planning system (TPS) and the same was executed in the accelerator and measured with the I'matriXX device. The dose distribution pattern measured by the matrix device for field-in-field, pyramidal, and chair test were found to be in good agreement with the calculated dose distribution by TPS both for 6 and 18 MV photons (gamma < or = 1: 96%, criteria 3%, 3 mm). Two 7-field IMRT plans (one prostate, one head and neck) dose distribution patterns were also measured with I'matriXX device and compared with film dosimetry. The measurements and evaluation proves that I'matriXX can be used for quantifying absolute dose. Moreover, using I'matriXX as absolute dosimeter in IMRT field verification, avoids the time-consuming procedure of making ionometric measurement for absolute dose estimation and film for dose distribution verification. The I'matriXX can also used for routine quality assurance checks like flatness, symmetry, field width, and penumbra of the linear accelerator beam.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol

Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...

متن کامل

Camera based EPID dosimetric verification of radiation treatments

Introduction: Tumor control and treatment quality depend on accuracy of delivered dose. Various methods have been studied for treatment dosimetry verification. Among these methods, the use of electronic portal imaging device (EPID) is highly considered. Many efforts are underway to improve the results of this context.   Materials and Methods: In this study, ...

متن کامل

Point Dose Measurement for Verification of Treatment Planning System using an Indigenous Heterogeneous Pelvis Phantom for Clarkson, Convolution, Superposition, and Fast Superposition Algorithms

Background: Nowadays, advanced radiotherapy equipment includes algorithms to calculate dose. The verification of the calculated doses is important to achieve accurate results. Mostly homogeneous dosimetric phantoms are available commercially which do not mimic the actual patient anatomy; therefore, an indigenous heterogeneous pelvic phantom mimicking actual human pelvic region has been used to ...

متن کامل

EPID in vivo Dosimetry

Introduction: The most modern radiotherapy devices are equipped with an Electronic Portal Imaging Device (EPID) system which is located on opposite side of the machine’s head. EPID system is often used to setting up the position verification during or between radiotherapy sessions. Material and Methods: Various types of dosimeters have been used to setting up ...

متن کامل

Characterization of a 2D ion chamber array for the verification of radiotherapy treatments.

In this study we investigated the characteristics of a commercial ion chamber array and its performance in the verification of radiotherapy plans. The device was the 2D Array Seven29 model (PTW, Freiburg, Germany). This is a two-dimensional detector array with 729 ionization chambers uniformly arranged in a 27 x 27 matrix with an active area of 27 x 27 cm(2). The detector short-, medium- and lo...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2010