Background: The reliability and validity of the SensorMedics VmaxST was tested. Methods: Thirty subjects (age = 24.5 ± 4.0 years, height = 174.8 ± 9.8 cm, weight = 70.3 ± 12.6 kg) performed treadmill exercise on three occasions, twice using the VmaxST and once using the SensorMedics 2900 system. Oxygen consumption (VO2; L/min) and heart rate (HR; beats/min) were measured continuously during three, 6minute stages: 80 m/min, 0% grade; 94 m/min, 5% grade; and 160 m/min, 0% grade, and VO2max. Results: Reliability was high, and measurement error was low for VO2 (Rxx range = 0.97 0.99, CI = 0.94 1.00, SEM = 0.03 0.08 L/min) and HR (Rxx = 0.94 0.99, CI = 0.88 1.00, SEM = 1.8 3.2 beats/min). Validity was high for VO2 (Rxy range = 0.92 0.98, CI = 0.84 0.99, SEE = 0.08 0.21 L/min) and HR (Rxy = 0.97 0.99, CI = 0.94 1.00, SEE = 0.9 1.8 beats/min). Mean differences in VO2 between VmaxST and 2900 were small yet significant (P < 0.001). Conclusions: The VmaxST demonstrated excellent reliability and validity for measuring VO2 and HR over several exercise intensities. Small overestimates in VO2 by the VmaxST are countered by low measurement error.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A method-comparison study regarding the validity and reliability of the Lactate Plus analyzer

OBJECTIVES The aims of this study were to: (1) determine the validity and reliability of the Nova Biomedical Lactate Plus portable analyzer, and quantify any fixed or proportional bias; (2) determine the effect of any bias on the determination of the lactate threshold and (3) determine the effect that blood sampling methods have on validity and reliability. DESIGN In this method comparison st...

متن کامل

Portable lactate analyzer for measuring lactate in cerebrospinal fluid (CSF) and plasma - method-comparison evaluations.

UNLABELLED Increased plasma lactate levels can indicate the presence of metabolic disorders in HIV infected individuals. OBJECTIVE To determine whether a portable analyzer is valid for measuring cerebrospinal fluid (CSF) and plasma lactate levels in HIV infected individuals. METHOD CSF and plasma were collected from 178 subjects. Samples tested by the Accutrend portable analyzer were compar...

متن کامل

Reliability of Body Landmarks Analyzer for Measuring the Quadriceps Angle

Genovarum and Genovalgum are the most common postural deformities of the knee joint. A quadriceps angle is used to measure these anomalies. Methods of measuring this angle are divided into two categories: invasive and non-invasive. The purpose of the present research was to study the inter/intra rater reliability of the non-invasive Body Landmarks Analyzer method for measuring of the quadriceps...

متن کامل

Validity and test-retest reliability of the Stride Analyzer in people with knee osteoarthritis.

INTRODUCTION Subjects with knee osteoarthritis walk differently compared to healthy subjects. Managing these gait alterations has been proven effective for reducing pain and increasing function. The Stride Analyzer is a low cost gait analysis tool but its clinimetric properties have not been investigated yet in subjects with symptomatic knee osteoarthritis. The aim of this study was to investig...

متن کامل

Monitoring Locomotor Load in Soccer: Is Metabolic Power, Powerful?

The aim of the present study was to examine the validity and reliability of metabolic power (P) estimated from locomotor demands during soccer-specific drills. 14 highly-trained soccer players performed a soccer-specific circuit with the ball (3×1-min bouts, interspersed with 30-s passive recovery) on 2 different occasions. Locomotor activity was monitored with 4-Hz GPSs, while oxygen update (V...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2004