Compliance Study of Endovascular Stent Grafts Incorporated with Polyester and Polyurethane Graft Materials in both Stented and Unstented Zones
نویسندگان
چکیده
Compliance mismatch between stent graft and host artery may induce complications and blood flow disorders. However, few studies have been reported on stent graft compliance. This study aims to explore the deformation and compliance of stent graft in stented and unstented zones under three pressure ranges. Compliance of two stent grafts incorporated with polyurethane graft (nitinol-PU) and polyester graft (nitinol-PET) materials respectively were tested; the stents used in the two stent grafts were identical. For the circumferential deformation of the stent grafts under each pressure range, the nitinol-PET stent graft was uniform in both zones. The nitinol-PU stent graft was circumferentially uniform in the stented zone, however, it was nonuniform in the unstented zone. The compliance of the PU graft material was 15 times higher than that of the PET graft. No significant difference in compliance was observed between stented and unstented zones of the nitinol-PET stent graft regardless of the applied pressure range. However, for the nitinol-PU stent graft, compliance of the unstented PU region was approximately twice that of the stented region; thus, compliance along the length of the nitinol-PU stent graft was not constant and different from that of the nitinol-PET stent graft.
منابع مشابه
Finite element analysis of the mechanical performances of 8 marketed aortic stent-grafts.
PURPOSE To assess numerically the flexibility and mechanical stresses undergone by stents and fabric of currently manufactured stent-grafts. METHODS Eight marketed stent-graft limbs (Aorfix, Anaconda, Endurant, Excluder, Talent, Zenith Flex, Zenith LP, and Zenith Spiral-Z) were modeled using finite element analysis. A numerical benchmark combining bending up to 180° and pressurization at 150 ...
متن کاملThe geometry of unstented and stented pig common carotid artery bypass grafts.
The long-term success of arterial bypass grafting with autologous saphenous veins is limited by neointimal hyperplasia (NIH), which seemingly develops preferentially at sites where hydrodynamic wall shear is low. Placement of a loose-fitting, porous stent around end-to-end, or end-to-side, autologous saphenous vein grafts on the porcine common carotid artery has been found significantly to redu...
متن کاملPii: S0967-2109(01)00120-x
10 years of experience with endovascular AAA repair has yielded important information regarding the relationship between stent-graft design and stent-graft performance. Commercially available systems differ in important ways. General conclusions regarding endovascular AAA repair need to be qualified by reference to a specific system, or systems. Nevertheless, one can draw conclusions regarding ...
متن کاملIncreased intimal hyperplasia in experimental vein graft stenting compared to arterial stenting: comparisons in a new rabbit model of stent injury.
BACKGROUND In-stent restenosis due to intimal hyperplasia is an important clinical problem. Animal models of stent injury are limited by inconsistent arterial responses to stenting, and less intimal hyperplasia than diseased human vessels. To address these issues, we aimed to compare the degree of intimal hyperplasia in stented rabbit jugular-carotid interposition grafts (vein grafts) versus st...
متن کاملInhibition of accelerated atherosclerosis in vein grafts by placement of external stent in apoE*3-Leiden transgenic mice.
OBJECTIVE Vein grafts fail because of the development of intimal hyperplasia and accelerated atherosclerosis. Placement of an external stent around vein grafts resulted in an inhibition of intimal hyperplasia in several animal studies. Here, we assess the effects of external stenting on accelerated atherosclerosis in early vein grafts in carotid arteries in hypercholesterolemic apolipoprotein E...
متن کامل