Internal mechanical conditions in the soft tissues of a residual limb of a trans-tibial amputee.

نویسندگان

  • S Portnoy
  • Z Yizhar
  • N Shabshin
  • Y Itzchak
  • A Kristal
  • Y Dotan-Marom
  • I Siev-Ner
  • A Gefen
چکیده

Most trans-tibial amputation (TTA) patients use a prosthesis to retain upright mobility capabilities. Unfortunately, interaction between the residual limb and the prosthetic socket causes elevated internal strains and stresses in the muscle and fat tissues in the residual limb, which may lead to deep tissue injury (DTI) and other complications. Presently, there is paucity of information on the mechanical conditions in the TTA residual limb during load-bearing. Accordingly, our aim was to characterize the mechanical conditions in the muscle flap of the residual limb of a TTA patient after donning the prosthetic socket and during load-bearing. Knowledge of internal mechanical conditions in the muscle flap can be used to identify the risk for DTI and improve the fitting of the prosthesis. We used a patient-specific modelling approach which involved an MRI scan, interface pressure measurements between the residual limb and the socket of the prosthesis and three-dimensional non-linear large-deformation finite-element (FE) modelling to quantify internal soft tissue strains and stresses in a female TTA patient during static load-bearing. Movement of the truncated tibia and fibula during load-bearing was measured by means of MRI and used as displacement boundary conditions for the FE model. Subsequently, we calculated the internal strains, strain energy density (SED) and stresses in the muscle flap under the truncated bones. Internal strains under the tibia peaked at 85%, 129% and 106% for compression, tension and shear strains, respectively. Internal strains under the fibula peaked at substantially lower values, that is, 19%, 22% and 19% for compression, tension and shear strains, respectively. Strain energy density peaked at the tibial end (104kJ/m(3)). The von Mises stresses peaked at 215kPa around the distal end of the tibia. Stresses under the fibula were at least one order of magnitude lower than the stresses under the tibia. We surmise that our present patient-specific modelling method is an important tool in understanding the etiology of DTI in the residual limbs of TTA patients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Internal Mechanical Conditions in the Soft Tissues of the Residual Limb in Trans- Tibial Amputees

Tibial Amputees 1Portnoy S., 2Shabshin N., 2Itzchak Y., 3Yizhar Z., 4Dotan-Marom Y., 4Oppenheim U., 4Siev-Ner I. and Gefen A. 1Biomedical Engineering, Tel Aviv University, Tel Aviv, Israel 2Diagnostic Imaging, Chaim Sheba Medical Center, Israel Physical Therapy, Tel Aviv University, Israel Orthopaedic Rehabilitation Department, Chaim Sheba Medical Center, Israel Web: www.eng.tau.ac.il/~msbm , e...

متن کامل

Investigation of lower-limb tissue perfusion during loading.

An extant tissue indentor used for amputee residual limb tissue indentation studies was modified to include laser Doppler flowmetry (LDF) to enable measurement of tissue perfusion during indentation. This device allows quantitative assessment of the mechanical and physiological response of soft tissues to load, as demonstrated by indentation studies of the lower-limb tissues of young healthy su...

متن کامل

Epithelialization Over a Scaffold of Antibiotic-Impregnated PMMA Beads: A Salvage Technique for Open Tibial Fractures with Bone and Soft Tissue Loss When all Else Fails

  The management of soft tissue defects in tibial fractures is essential for limb preservation. Current techniques are not without complications and may lead to poor functional outcomes. A salvage method is described using three illustrative cases whereby a combination of flaps and antibiotic-impregnated polymethylmethacrylate beads are employed to fill the bony defect, fight the infection, and...

متن کامل

Computer-aided design of a prosthetic socket for an above-knee amputee.

A computer-aided design process for fabricating the rectified cast for an above-knee prosthetic socket is described. The methodology for collecting the parameters required for the computer analysis is discussed. The input variables include the unloaded shape of the residual limb, the mechanical properties of the soft tissues that comprise the limb, and the surface loading that deforms the tissu...

متن کامل

Joint loading and bone mineral density in persons with unilateral, trans-tibial amputation.

BACKGROUND Persons with unilateral, lower-extremity amputation sometimes develop osteoarthritis in the intact limb. The purpose of this study was to investigate gait mechanics and bone mineral density in unilateral, trans-tibial amputees to test the hypotheses that the intact limb knee and hip will have larger frontal plane net joint moments and bone mineral density than the prosthetic side and...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of biomechanics

دوره 41 9  شماره 

صفحات  -

تاریخ انتشار 2008