Assessment of phantom dosimetry and image quality of i-CAT FLX cone-beam computed tomography.

نویسندگان

  • John B Ludlow
  • Cameron Walker
چکیده

INTRODUCTION The increasing use of cone-beam computed tomography in orthodontics has been coupled with heightened concern about the long-term risks of x-ray exposure in orthodontic populations. An industry response to this has been to offer low-exposure alternative scanning options in newer cone-beam computed tomography models. METHODS Effective doses resulting from various combinations of field of view size and field location comparing child and adult anthropomorphic phantoms with the recently introduced i-CAT FLX cone-beam computed tomography unit (Imaging Sciences, Hatfield, Pa) were measured with optical stimulated dosimetry using previously validated protocols. Scan protocols included high resolution (360° rotation, 600 image frames, 120 kV[p], 5 mA, 7.4 seconds), standard (360°, 300 frames, 120 kV[p], 5 mA, 3.7 seconds), QuickScan (180°, 160 frames, 120 kV[p], 5 mA, 2 seconds), and QuickScan+ (180°, 160 frames, 90 kV[p], 3 mA, 2 seconds). Contrast-to-noise ratio was calculated as a quantitative measure of image quality for the various exposure options using the QUART DVT phantom. RESULTS Child phantom doses were on average 36% greater than adult phantom doses. QuickScan+ protocols resulted in significantly lower doses than standard protocols for the child (P = 0.0167) and adult (P = 0.0055) phantoms. The 13 × 16-cm cephalometric fields of view ranged from 11 to 85 μSv in the adult phantom and 18 to 120 μSv in the child phantom for the QuickScan+ and standard protocols, respectively. The contrast-to-noise ratio was reduced by approximately two thirds when comparing QuickScan+ with standard exposure parameters. CONCLUSIONS QuickScan+ effective doses are comparable with conventional panoramic examinations. Significant dose reductions are accompanied by significant reductions in image quality. However, this trade-off might be acceptable for certain diagnostic tasks such as interim assessment of treatment results.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

History of imaging in orthodontics: A history of overdiagnosis?

4. Radiation protection: cone beam CT for dental and maxillofacial radiology. Evidence based guidelines. 2011. Available at: http:// www.sedentexct.eu/files/guidelines_final.pdf. Accessed December 23, 2015. 5. Halazonetis DJ. Cone-beam computed tomography is not the imaging technique of choice for comprehensive orthodontic assessment [Point/Counterpoint] Am J Orthod Dentofacial Orthop 2012; 141...

متن کامل

Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography

Background: While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free.Objective: The aim of this study was to evaluate the exposure doses received by patients during maxillofacial imaging with CBCT.Methods: Entrance surface dose (ESD) was measured by using thermoluminescent dosimeters (TLDs) attached to the eyes...

متن کامل

Usability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter

Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...

متن کامل

Skin Doses on the Lens for Temporomandibular Joint Exam in Cone Beam Computed Tomography

The aim of this study was to evaluate the kerma at the surface of the lens in TMJ CBCT and to derive the equivalent dose. An anthropomorphic phantom of the head and neck manufactured by Radiation Support Devices (model RS230) was used. The dosimetric measurements were obtained by using fourteen thermoluminescent dosimetry (TLD) dosimeters (LiF: Mg, Ti), divided in two pairs (one pair for each e...

متن کامل

Artifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality

Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • American journal of orthodontics and dentofacial orthopedics : official publication of the American Association of Orthodontists, its constituent societies, and the American Board of Orthodontics

دوره 144 6  شماره 

صفحات  -

تاریخ انتشار 2013