Analysis of complex framing coil stability in a wide-necked aneurysm model.
نویسندگان
چکیده
Appropriately sized 0.010- and 0.018-inch complex framing coils were placed in a wide-necked silicone aneurysm replica, and their stability was evaluated at variable physiologic flow rates using video recording. After detachment, the 0.010-inch coils demonstrated instability/prolapse that was proportional to flow rate. In contrast, 0.018-inch coils held their 3D configuration regardless of flow rate. The findings support the use of 0.018-inch coils (when possible) in aneurysms with unfavorable geometry, particularly in circulations with higher flow rates.
منابع مشابه
Transluminal stent-assisted coil embolization of a vertebral confluence aneurysm: technique report.
BACKGROUND Recent advances in stent technology have allowed for negotiation of often tortuous posterior circulation intracranial vasculature. Stent-assisted coil embolization is a novel treatment for complex wide-necked aneurysms, as stents provide a buttress that allows for coil deposition while preventing coil herniation into the parent vessel lumen. We describe a case of stent-assisted coil ...
متن کاملSingle-center experience in the endovascular treatment of wide-necked intracranial aneurysms with a bridging intra-/extra-aneurysm implant (pCONus)
PURPOSE To retrospectively evaluate the safety and efficacy of the endovascular treatment of wide-necked intracranial aneurysms assisted by a novel intra-/extra-aneurysm stent-like implant (pCONus). METHODS Initial and follow-up angiographic and clinical results are presented of 25 patients with 25 unruptured and ruptured wide-necked intracranial aneurysms treated by reconstruction of the ane...
متن کاملUse of Triple Microcatheters for Endovascular Treatment of Wide-Necked Intracranial Aneurysms: A Single Center Experience
OBJECTIVE The dual microcatheter technique is common practice for coil embolization of a wide-necked aneurysm, due to safety and efficacy. However, technical limitations of some complex configurations may necessitate additional microcatheters to bolster coil stability, compact the coil, or for protection. Described herein is a triple microcatheter technique for endovascular management of wide-n...
متن کاملUse of three-dimensional Guglielmi detachable coils in the treatment of wide-necked cerebral aneurysms.
A three-dimensional Guglielmi detachable coil (3D-GDC) has been developed that is designed specifically to bridge the neck of an aneurysm with coil loops, thereby facilitating retention of additional coils placed within the aneurysm. Nine wide-necked cerebral aneurysms were successfully embolized using the 3D-GDC, provided that the dome-to-neck ratio was 1.5 or greater.
متن کاملCombined stent implantation and endosaccular coil placement for treatment of experimental wide-necked aneurysms: a feasibility study in swine.
PURPOSE To assess the feasibility of combining stent implantation in the parent artery with endosaccular coil placement for the treatment of experimentally created wide-necked aneurysms. METHODS Wide-necked aneurysms were surgically created on the common carotid artery in 12 swine. A metal stent was endovascularly implanted across each aneurysm neck and its effect documented anigiographically...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- AJNR. American journal of neuroradiology
دوره 28 2 شماره
صفحات -
تاریخ انتشار 2007