Efficiency of manual resuscitators at birth.

نویسندگان

  • D Field
  • A D Milner
  • I E Hopkin
چکیده

The effectiveness of five neonatal/paediatric manual resuscitators was assessed in a group of babies born by caesarean section. Results showed that devices incorporating a large volume reservoir produced the greatest tidal volume, while those with smaller volume reservoirs could not be considered satisfactory for routine use during neonatal resuscitation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of 16 adult disposable manual resuscitators.

INTRODUCTION Disposable manual resuscitators are commonly used to ventilate patients during cardiopulmonary resuscitation, suctioning, and intrahospital transport, and their clinical performance is critical. METHODS We bench-tested 16 adult disposable manual resuscitators from 9 different manufacturers. We performed a series of tests and made observations using testing industry standards as a...

متن کامل

Evaluation of manual resuscitators used in ICUs in Brazil*

OBJECTIVE To evaluate the performance of manual resuscitators (MRs) used in Brazil in accordance with international standards. METHODS Using a respiratory system simulator, four volunteer physiotherapists employed eight MRs (five produced in Brazil and three produced abroad), which were tested for inspiratory and expiratory resistance of the patient valve; functioning of the pressure-limiting...

متن کامل

Manual Resuscitators: Some Inconvenient Truths

The Old Testament appears to describe the successful application of mouth-to-mouth resuscitation of a child, some 2,900 years ago (II Kings 4:34). According to O’Donnell and colleagues,1 the Babylonian Talmud describes a similar event 2,300 years ago: “One may give a woman (about to give birth) all assistance possible.... One may violate the Sabbath on her account.” What is meant by “being of a...

متن کامل

Comparison of the FiO2 delivered by seven models of the self-inflating bag-mask system.

BACKGROUND AND OBJECTIVES Since resuscitators with self-inflating bag-mask systems manufactured and/or commercialized in Brazil are widely available and used in health services, both out- and intra-hospitals, the objective of this study was to determine the O2 fractions delivered by seven resuscitators receiving different O2 flows. METHODS Seven resuscitators with self-inflating bag-mask syst...

متن کامل

[Failure of self-inflating manual resuscitator due to the presence of dried pulmonary secretions].

BACKGROUND AND OBJECTIVES Self-inflating manual resuscitators (SIMR) are devices used to ventilate patients with ventilatory needs. These devices consist of a set of valves that work sequentially, and changes in their function may be harmful to patients. CASE REPORT During the use of SIMR, it was observed that the valve remained fixed due to the presence of dried pulmonary secretion, which ma...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Archives of disease in childhood

دوره 61 3  شماره 

صفحات  -

تاریخ انتشار 1986