Bias and precision of cholesterol analysis by physician's office analyzers.

نویسندگان

  • T R Koch
  • U Mehta
  • H Lee
  • K Aziz
  • S Temel
  • J A Donlon
  • R Sherwin
چکیده

We studied the bias and precision of serum cholesterol analysis by physician's office analyzers. Total imprecision (CV range, %) for analysis of serum pools was: Abbott Vision 1.5%-1.9%; Ames Seralyzer 3.9%-4.5%; BMD Reflotron 2.3%-3.8%; Chrometrics Cholesterol Test System 2.3%-2.8%; Kodak DT-60 1.6%-2.2%. The Ames Seralyzer exhibited an excessive between-run component of variation. We collected, from 109 volunteers, samples of venous serum, heparin-treated whole blood, heparin-treated plasma, and fingerstick whole blood, and analyzed each type (where possible) with each system; serum was also analyzed in duplicate (by a proposed Reference Method) at the Centers for Disease Control. For assays with serum, the BMD Reflotron and Kodak DT-60 exhibited negative bias. All systems gave lower results for plasma and whole blood than for serum. All systems except the Kodak DT-60 were less precise for analysis of patients' sera than for analysis of serum pools; between-specimen variables may influence the results of these systems.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Comparative Study on Instrumental Precision of Refrigerated and Non-Refrigerated Auto-Analyzers in Order to Improve Quality Assurance in Biochemistry Laboratory

Background and Objective: Quality control is one of the most important components in order to improve quality assurance in laboratories during analytical steps. For this purpose, coefficient of variation plays an important role. Due to the fast improvement in technology, application of inferential statistics for the comparisons ...

متن کامل

Another physician's office analyzer: the Abbott "Vision" evaluated.

We evaluated the Abbott Vision system, to assess its suitability for use in the physician's office setting. We compared results from Vision for all the available analytes (alkaline phosphatase, urea nitrogen, cholesterol, creatinine, glucose, uric acid, and triglycerides) with results from the Kodak Ektachem 400 and 700. For all analytes, standard curve linearity, assessed with various dilution...

متن کامل

Laboratory and field evaluation of three desktop instruments for assay of cholesterol and triglyceride.

We evaluated three desktop instruments suitable for decentralizing the assays of blood cholesterol and triglycerides to satellite and physician's office testing. The instruments, Ektachem DT 60 (E), Reflotron (R), and Seralyzer (S), were used according to their manufacturers' instructions to assay serum or capillary blood from outpatients at two physicians offices. Accuracy was assessed in the ...

متن کامل

Inaccuracy of lipid measurements with the portable Cholestech L.D.X analyzer in patients with hypercholesterolemia.

BACKGROUND Although total cholesterol concentrations measured by portable lipid analyzers have acceptable bias and precision in young and middle-aged adults, clinically relevant differences in HDL-cholesterol (HDL-C) and triglyceride values have been described. Furthermore, the accuracy of portable lipid analyzers in older hyperlipidemic individuals, who have a high incidence of coronary heart ...

متن کامل

Evaluation of seven Cholestech L.D.X analyzers for total cholesterol determinations.

We assessed the performance of seven Cholestech L.D.X lipid analyzers under tightly controlled laboratory conditions for accuracy and precision in accordance with analytical guidelines of the National Cholesterol Education Program (NCEP). Venous heparinized whole blood (VB) and plasma (VP), venous serum (VS), and capillary fingerstick whole blood (FB) were collected from 18 individuals. Total c...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 33 12  شماره 

صفحات  -

تاریخ انتشار 1987