Positive interference by danazol in a testosterone radioimmunoassay kit procedure.

نویسندگان

  • A M Sharp
  • I S Fraser
  • S Robertson
  • J R Turtle
چکیده

We describe a problem in measuring serum testosterone concentrations in the serum of women who are taking danazol. The Diagnostic Products Corporation testosterone radioimmunoassay gives very high apparent testosterone concentrations in these women and the value obtained depends on the volume of serum extracted per assay tube. Chromatography of the serum extracts on Sephadex LH-20, but not on Celite, before radioimmunoassay reduced these high apparent testosterone concentrations to less than that found in normal women.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Further studies on danazol interference in testosterone radioimmunoassays.

Investigating the interference of danazol in testosterone radioimmunoassays, we measured serum testosterone concentrations in the serum of women receiving danazol, with three testosterone radioimmunoassay kits. All of these showed positive interference by danazol or its metabolites (or both). Values for testosterone obtained with two of the methods (Diagnostic Products Corp. and Farmos Diagnost...

متن کامل

Positive interference in Radioimmunoassay determination of serum total T3

Introduction: Determination of thyrotropin, total and free thyroxine and triiodothyronine are widely used as diagnostic methods for thyroid function evaluation. There have been numerous reports of interferences in thyroid hormone immunoassays. Herein, a prominent positive interference is reported for a radioimmunoassay kit of total triiodothyronine. Materials and Methods: In this study, a total...

متن کامل

Direct measurement of testosterone in a pediatric center, with use of a radioimmunoassay kit and unextracted serum.

We evaluated a commercial (Radioassay Systems) radioimmunoassay kit for testosterone in which a double-antibody method and an 125I tracer are used. The 50-microL serum sample is assayed directly with no pre-extraction step. The assay system has good sensitivity (0.01 microgram/L), good precision (CV 5.2%), and little cross reactivity between testosterone and other androgens. Results compared we...

متن کامل

Total testosterone, free-androgen index, calculated free testosterone, and free testosterone by analog RIA compared in hirsute women and in otherwise-normal women with altered binding of sex-hormone-binding globulin.

We compared the clinical value of information on free testosterone as measured with the Coat-A-Count (Diagnostic Products Corp.) radioimmunoassay kit involving a ligand analog with that of total testosterone, the free-androgen index, and free testosterone calculated from concentrations of testosterone, sex-hormone-binding globulin, and albumin, in hirsute women, pregnant women, oral-contracepti...

متن کامل

Local Preparation and Evaluation of Double-antibody Liquid Phase Radioimmunoassay System for Detection of Human Testosterone

Preparation, evaluation and optimization of testosterone radioimmunoassay (RIA) system using liquid phase double antibody is considered to be the main objective. Three primary components were prepared and characterized to obtain valid and accurate system. These components were polyclonal testosterone antibody, the I-testosterone tracer and set of testosterone standards. The production of polycl...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 27 4  شماره 

صفحات  -

تاریخ انتشار 1981