The added diagnostic value of five different C-reactive protein point-of-care test devices in detecting pneumonia in primary care: A nested case-control study.
نویسندگان
چکیده
BACKGROUND The results obtained from various point-of-care (POC) test devices for estimating C-reactive protein (CRP) levels in a laboratory setting differ when compared to a laboratory reference test. We aimed to determine whether such differences meaningfully affect the accuracy and added diagnostic value in predicting radiographic pneumonia in adults presenting with acute cough in primary care. METHODS A nested case control study of adult patients presenting with acute cough in 12 different European countries (the Genomics to combat Resistance against Antibiotics in Community-acquired LRTI in Europe [GRACE] Network). Venous blood samples from 100 patients with and 100 patients without pneumonia were tested with five different POC CRP tests and a laboratory analyzer. Single test accuracy values and the added value of CRP to symptoms and signs were calculated. RESULTS Single test accuracy values showed similar results for all five POC CRP tests and the laboratory analyzer. The area under the curve of the different POC CRP tests and the laboratory analyzer (range 0.79-0.80) were all comparable and higher than the clinical model without CRP (0.70). Multivariable odds ratios were the same (1.2) for all CRP tests. CONCLUSIONS Five POC CRP test devices and the laboratory analyzer performed with similar accuracy in detecting pneumonia both as single test, and when used in addition to clinical findings. Variability in results obtained from standard CRP laboratory and POC test devices do not translate into clinically relevant differences when used for prediction of pneumonia in patients with acute cough in primary care.
منابع مشابه
Diagnostic value of urinary antigen Streptococcus pneumoniae in children With pneumonia: A case control study
Abstract Background and objectives: The aim of this study was the detection of S.pneumoniae infection by rapid urinary test and blood culture in children with pneumonia in comparison with healthy children. Material and Methods: This case control study was carried out in pediatric ward of Rasoul Akram hospital in Tehran, Iran (2006 - 200٧).Fifty-four Community acquired pneumonia (CAP) and 50 hea...
متن کاملIdentifying high and low serum-ascites albumin gradient in ascitic fluid by the point of care dipstick test
Objective: To evaluate the capability of ascitic fluid dipstick results for pH, glucose, and protein in order to predict a low serum-ascites albumin gradient (SAAG) at the bedside of the patient in the emergency department (ED). Methods: This prospective cross-sectional study was conducted during one year in the ED of Afzalipour hospital in Kerman, Iran. ...
متن کاملDiagnosing serious infections in acutely ill children in ambulatory care (ERNIE 2 study protocol, part A): diagnostic accuracy of a clinical decision tree and added value of a point-of-care C-reactive protein test and oxygen saturation
BACKGROUND Acute illness is the most common presentation of children to ambulatory care. In contrast, serious infections are rare and often present at an early stage. To avoid complications or death, early recognition and adequate referral are essential. In a recent large study children were included prospectively to construct a symptom-based decision tree with a sensitivity and negative predic...
متن کاملارزش تشخیصی لاکتات دهیدروژناز و کراتینین مایع مهبلی در تشخیص پارگی زودرس پرده های جنینی
Background and purpose: Premature rupture of membrane (PRÔM) is a common and challenging issue in obstetrics. Ëarly diagnosis of PRÔM and the consequent infection has a pivotal role in proper management of patients. The detection of PRÔM is easier in a case with visible vaginal fluid, but is difficult when trivial amount of vaginal fluid presents. Ïn this study, we aimed to assess the diagnos...
متن کاملSignificance of Serum Procalcitonin Level in the Early Diagnosis of Neonatal Sepsis
Background: Sepsis in neonates presents itself with non-specific clinical features which makes early diagnosis difficult. However, procalcitonin (PCT) and other inflammatory markers have recently been considered as sensitive markers for the early detection of neonatal sepsis. Therefore, the present study aimed to determine the diagnostic value of PCT in the early detection of neonatal sepsis an...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Scandinavian journal of clinical and laboratory investigation
دوره 75 4 شماره
صفحات -
تاریخ انتشار 2015