Evaluation of the automated phoenix system for potential routine use in the clinical microbiology laboratory.

نویسندگان

  • J-L Donay
  • D Mathieu
  • P Fernandes
  • C Prégermain
  • P Bruel
  • A Wargnier
  • I Casin
  • F X Weill
  • P H Lagrange
  • J L Herrmann
چکیده

A comparative study was designed to evaluate the identification (ID) and antimicrobial susceptibility testing (AST) performances of the BD Phoenix Automated Microbiology System (Becton Dickinson Diagnostic Systems [BD], Pont de Claix, France). A total of 305 single clinical isolates were collected, and comparisons were made with routine manual methods in use in our microbiology laboratories. The percentages of correct IDs were 93.3, 89.4, 91.8, and 85.7% for enterobacteria, nonfermenting gram-negative bacilli, staphylococci, and streptococci-enterococci, respectively. The median ID time was 3 h, and the median time for AST was 10 h 30 min. AST results showed variable percentages of errors for the different antibiotics. None of the enterobacteria and 0.3% of Pseudomonas aeruginosa isolates showed a very major error (VME). Only one strain of Staphylococcus aureus showed a VME with oxacillin. We demonstrate here the efficiency of the Phoenix system, which can be used for the majority of strains encountered in a university-based laboratory, for ID and AST.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of an Automated System with Conventional Identification and Antimicrobial Susceptibility Testing

The present study was designed to compare a fully automated identification/antibiotic susceptibility testing (AST) system BD Phoenix (BD) for its efficacy in rapid and accurate identification and AST with conventional manual methods and to determine if the errors reported in AST, such as the (very major errors) VME (false susceptibility), (major errors) ME (false resistance), and (minor errors)...

متن کامل

Accuracies of beta-lactam susceptibility test results for Pseudomonas aeruginosa with four automated systems (BD Phoenix, MicroScan WalkAway, Vitek, and Vitek 2).

Contemporary clinical isolates and challenge strains of Pseudomonas aeruginosa were tested by four automated susceptibility testing systems (BD Phoenix, MicroScan WalkAway, Vitek, and Vitek 2; two laboratories with each) against six broad-spectrum beta-lactams, and the results were compared to reference broth microdilution (BMD) and to consensus results from three validated methods (BMD, Etest ...

متن کامل

Evaluation of the Etest ESBL and the BD Phoenix, VITEK 1, and VITEK 2 automated instruments for detection of extended-spectrum beta-lactamases in multiresistant Escherichia coli and Klebsiella spp.

Seventy-four isolates of multiresistant Escherichia coli and Klebsiella spp. recovered during a 3-year period and 17 control strains with genotypically identified beta-lactamases were tested for the production of extended-spectrum beta-lactamases (ESBLs) by using the Etest and the VITEK 1, VITEK 2, and Phoenix automated instruments. The use of the Etest was evaluated by investigating its accura...

متن کامل

Two-center collaborative evaluation of the performance of the BD Phoenix automated microbiology system for identification and antimicrobial susceptibility testing of Enterococcus spp. and Staphylococcus spp.

The performance of the BD Phoenix Automated Microbiology System (BD Diagnostic Systems, Sparks, Md.) was assessed for identification (ID) and antimicrobial susceptibility testing (AST) for the majority of clinically encountered bacterial isolates in a European collaborative two-center trial. A total of 469 bacterial isolates of the genera Staphylococcus (275 isolates), Enterococcus (179 isolate...

متن کامل

Evaluation of vancomycin susceptibility testing for methicillin-resistant Staphylococcus aureus: comparison of Etest and three automated testing methods.

We evaluated the ability of four commercial MIC testing systems (MicroScan, Vitek 2, Phoenix, and Etest) to detect vancomycin MIC values of ≤1 to ≥2 in 200 methicillin-resistant Staphylococcus aureus (MRSA) strains compared to the Clinical and Laboratory Standards Institute broth microdilution (BMD) reference methods. Compared to the BMD method, absolute agreement (0 ± dilution) was highest for...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 42 4  شماره 

صفحات  -

تاریخ انتشار 2004