Repeatability and Reproducibility of Macular Thickness Measurements Using Fourier Domain Optical Coherence Tomography
نویسندگان
چکیده
AIM To evaluate repeatability and reproducibility of macular thickness measurements in visually normal eyes using the Topcon 3D OCT-1000. METHODS Phase 1 investigated scan repeatability, the effect of age and pupil dilation. Two groups (6 younger and 6 older participants) had one eye scanned 5 times pre and post- dilation by 1 operator. Phase 2 investigated between-operator, within and between-visit reproducibility. 10 participants had 1 un-dilated eye scanned 3 times on 2 separate visits by 2 operators. RESULTS PHASE 1: No significant difference existed between repeat scans (p=0.75) and no significant difference was found pre- and post-dilation (p=0.54). In the younger group variation was low (95% limits +/- 3.62microm) and comparable across all retinal regions. The older group demonstrated greater variation (95% limits +/- 7.6microm). PHASE 2: For a given retinal location, 95% confidence limits for within-operator, within-visit reproducibility was 5.16microm. This value increased to 5.56microm for the same operator over two visits and to 6.18microm for two operators over two visits. CONCLUSION A high level repeatability, close to 6microm, of macular thickness measurement is possible using the 3D OCT- 1000. Measured differences in macular thickness between successive visits that exceed 6microm in pre-presbyopic individuals are therefore likely to reflect actual structural change. OCT measures are more variable in older individuals and it is advisable to take a series of scans so that outliers can be more easily identified.
منابع مشابه
بررسی پایایی و تکرار پذیری دستگاه Spectral domain optical coherence tomography در اندازه گیری ضخامت ماکولا در چشمهای طبیعی قبل و بعد از گشاد کردن مردمک
Background: optical coherence tomography is one of the most valuable imaging techniques in the evaluation of macula. Objective: to investigate the repeatability and reproducibility of OCT in measuring the macular thickness in normal eyes before and after pupil dilation. Methods: A total of 44 eyes were enrolled. All subjects underwent macular thickness measurement using the Cirrus SD-OCT ma...
متن کاملRetinal Ganglion Cell Complex in Alzheimer Disease: Comparing Ganglion Cell Complex and Central Macular Thickness in Alzheimer Disease and Healthy Subjects Using Spectral Domain-Optical Coherence Tomography
Introduction: Alzheimer disease (AD) is the most common form of dementia worldwide. The modalities to diagnose AD are generally expensive and limited. Both the central nervous system (CNS) and the retina are derived from the cranial neural crest; therefore, changes in retinal layers may reflect changes in the CNS tissue. Optical coherence tomography (OCT) machine can show delicate retinal layer...
متن کاملReproducibility of retinal thickness measurements in patients with age-related macular degeneration using 3D Fourier-domain optical coherence tomography (OCT) (Topcon 3D-OCT 1000).
PURPOSE Conventional time-domain optical coherence tomography (OCT) has become an important tool for following dry or exudative age-related macular degeneration (AMD). Fourier-domain three-dimensional (3D) OCT was recently introduced. This study tested the reproducibility of 3D-OCT retinal thickness measurements in patients with dry and exudative AMD. METHODS Ten eyes with dry AMD and 12 eyes...
متن کاملRepeatability and reproducibility of fast macular thickness mapping with stratus optical coherence tomography.
OBJECTIVE To determine the repeatability and reproducibility of retinal thickness measurements with the fast macular thickness mapping protocol of Stratus optical coherence tomography. METHODS Ten eyes of 10 healthy subjects and 15 eyes of 15 diabetic patients with clinically significant macular edema (CSME) underwent 2 scanning sessions before and after pupil dilation during the same visit b...
متن کاملComparison of Retinal Nerve Fiber Layer Thickness Measurements in Healthy Subjects Using Fourier and Time Domain Optical Coherence Tomography
Purpose. To compare the retinal nerve fiber layer (RNFL) measurements using two different ocular coherence tomography (OCT) devices: Cirrus Fourier domain OCT and Stratus time domain OCT. To analyze reproducibility of Fourier domain measurements in healthy subjects. Methods. One hundred and thirty-two eyes of 132 healthy subjects were scaned on the same day with both instruments, separated by 1...
متن کامل