Refinement of Intraocular Pressure Measurements made by Ocular Response Analyzer following Laser in Situ Keratomileusis using M2 90 Moria Microkeratome for Egyptian Myopic and Astigmatic Patients
نویسندگان
چکیده
INTRODUCTION The Goldmann applanation tonometer readings were noted to be markedly reduced after Lasik surgery using a thin flap technique which is widely used nowadays, to correct a wide range of myopia and astigmatism. The Ocular Response Analyzer (ORA) is considered one of the important tools for refinement of Intraocular Pressure (IOP) measurements. The aim of this study is to evaluate and refine ORA measurements for IOP changes post- M2 90 Moria microkeratome Lasik procedure for Egyptian myopic and astigmatic patients trying to aid in glaucoma diagnosis and management. METHODS Thirty-five eyes of nineteen Egyptian patients with myopia or myopic astigmatism who had undergone Lasik procedure using M2 90 Moria microkeratome were included in this consecutive case series study. All cases were subjected to full ophthalmological examination including uncorrected and best corrected visual acuity, refraction, slit lamp examination, Scheimpflug imaging and ORA (Reichert Ophthalmic Instruments, Inc., Buffalo, NY, USA). The refractive errors were measured using an auto refractometer and were confirmed by trial. For the ocular response analyzer, the Goldmann-correlated IOP measurement (IOPg) which simulates IOP measured by Goldmann tonometer and the Corneal-Compensated Intraocular Pressure (IOPcc) that takes corneal biomechanical properties into consideration were reported as well as the Corneal Hysteresis (CH) and the Corneal Resistance Factor (CRF). Cases were examined preoperatively and again about one month after Lasik, after stabilization of eye condition. The collected data were tabulated and analysed with the suitable statistical methods. The mean values and standard deviation were calculated for quantitative data. Comparison tests (t-test) and correlation tests (Pearson) were also performed. RESULTS In our study, involving M2 90 Moria microkeratome Lasik procedure, a highly significant post-Lasik reduction in IOPg (t-test = 8.62 (p<0.01), and a statistically significant reduction in IOPcc (t-test = 3.37 (p<0.05) by ocular response analyzer. The average post-Lasik reduction in IOP in mmHg was 4.84±2.82 for IOPg and 2.23±3.51for IOPcc. A statistically significant correlation (p<0.05) existed between post-Lasik reduction of IOPg and ablation depth as well as preoperative spherical equivalent. A non-significant correlation existed between post-Lasik reduction of IOPcc and both of ablation depth and preoperative spherical equivalent. CONCLUSIONS Following Lasik procedure using M2 90 Moria microkeratome, the IOPcc values by ORA were closer to the preoperative values than the IOPg values and the reduction of IOPcc values was not affected by ablation depth (i.e. preoperative spherical equivalent). Therefore, it is advisable to use the IOPcc values by ORA when evaluating post-Lasik patients for glaucoma diagnosis and management, but with the addition of the average value for post-Lasik reduction which in the study was 2.23±3.51mmHg.
منابع مشابه
The effect of flap thickness on corneal biomechanics after myopic laser in situ keratomileusis using the M-2 microkeratome
Purpose The purpose of this study was to determine the effect of flap thickness on corneal biomechanics after myopic laser in situ keratomileusis (LASIK). Methods This is a prospective controlled non-randomized, institutional study. Patients underwent either epi-LASIK with mitomycin (advanced surface ablation [ASA]), thin-flap LASIK (90 µm head), or thick-flap LASIK (130 µm head). In ASA, the...
متن کاملCorneal flap thickness with the Moria M2 single-use head 90 microkeratome in 72 consecutive LASIK procedures
PURPOSE This study aimed to evaluate the accuracy and consistency of corneal flap thickness in laser-assisted in situ keratomileusis (LASIK) with the Moria M2 single-use head 90 microkeratome. METHODS The central corneal thickness of 72 (37 right, 35 left) eyes of 37 patients was measured by ultrasonic pachymetry preoperatively and intraoperatively after flap cut. The Moria M2 single-use head...
متن کاملCorneal flap thickness with the Moria M2 single-use head 90 microkeratome.
PURPOSE To analyse the accuracy of corneal flap thickness created in laser-assisted in situ keratomileusis (LASIK) using the Moria Model 2 (M2) single-use head 90 microkeratome. METHODS The corneal thickness of 300 (266 myopic and 34 hyperopic) eyes of 150 patients was measured by ultrasonic pachymetry preoperationally and intraoperationally after flap cut. The Moria M2 single-use head 90, in...
متن کاملPorcine model to compare real-time intraocular pressure during LASIK with a mechanical microkeratome and femtosecond laser.
PURPOSE To compare real-time intraocular pressure (IOP) during laser in situ keratomileusis (LASIK) in porcine eyes using two types of microkeratomes. METHODS An interventional, prospective study of two microkeratomes: a Moria 2 (Moria group) and an IntraLase femtosecond laser (IntraLase Corp., Irvine, CA; IntraLase group). These devices were used to create lamellar corneal flaps in freshly e...
متن کاملQuality assessment model of 3 different microkeratomes through confocal microscopy.
PURPOSE To study the quality of the cut created by 3 microkeratomes from 2 different generations using corneal confocal microscopy. SETTING Department of Refractive Surgery, Instituto Oftalmológico de Alicante, Alicante, Spain. METHODS Two different studies were conducted: Study 1 and Study 2. Study 1 was a prospective analysis using confocal microscopy examination data from 2 reference gro...