Dynamic Stabilization with Active Locking Plates Delivers Faster, Stronger, and More Symmetric Fracture-Healing.
نویسندگان
چکیده
BACKGROUND Axial dynamization of fractures can promote healing, and overly stiff fixation can suppress healing. A novel technology, termed active plating, provides controlled axial dynamization by the elastic suspension of locking holes within the plate. This prospective, controlled animal study evaluated the effect of active plates on fracture-healing in an established ovine osteotomy model. We hypothesized that symmetric axial dynamization with active plates stimulates circumferential callus and delivers faster and stronger healing relative to standard locking plates. METHODS Twelve sheep were randomly assigned to receive a standard locking plate or an active locking plate for stabilization of a 3-mm tibial osteotomy gap. The only difference between plates was that locking holes of active plates were elastically suspended, allowing up to 1.5 mm of axial motion at the fracture. Fracture-healing was analyzed weekly on radiographs. After sacrifice at nine weeks postoperatively, callus volume and distribution were assessed by computed tomography. Finally, to determine their strength, healed tibiae and contralateral tibiae were tested in torsion until failure. RESULTS At each follow-up, the active locking plate group had more callus (p < 0.001) than the standard locking plate group. At postoperative week 6, all active locking plate group specimens had bridging callus at the three visible cortices. In standard locking plate group specimens, only 50% of these cortices had bridged. Computed tomography demonstrated that all active locking plate group specimens and one of the six standard locking plate group specimens had developed circumferential callus. Torsion tests after plate removal demonstrated that active locking plate group specimens recovered 81% of their native strength and were 399% stronger than standard locking plate group specimens (p < 0.001), which had recovered only 17% of their native strength. All active locking plate group specimens failed by spiral fracture outside the callus zone, but standard locking plate group specimens fractured through the osteotomy gap. CONCLUSIONS Symmetric axial dynamization with active locking plates stimulates circumferential callus and yields faster and stronger healing than standard locking plates. CLINICAL RELEVANCE The stimulatory effect of controlled motion on fracture-healing by active locking plates has the potential to reduce healing complications and to shorten the time to return to function.
منابع مشابه
Dynamic Stabilization of Simple Fractures With Active Plates Delivers Stronger Healing Than Conventional Compression Plating
OBJECTIVES Active plates dynamize a fracture by elastic suspension of screw holes within the plate. We hypothesized that dynamic stabilization with active plates delivers stronger healing relative to standard compression plating. METHODS Twelve sheep were randomized to receive either a standard compression plate (CP) or an active plate (ACTIVE) for stabilization of an anatomically reduced tib...
متن کاملComparison of 4 Methods for Dynamization of Locking Plates: Differences in the Amount and Type of Fracture Motion
BACKGROUND Decreasing the stiffness of locked plating constructs can promote natural fracture healing by controlled dynamization of the fracture. This biomechanical study compared the effect of 4 different stiffness reduction methods on interfragmentary motion by measuring axial motion and shear motion at the fracture site. METHODS Distal femur locking plates were applied to bridge a metadiap...
متن کاملStabilization of distal femur fractures with intramedullary nails and locking plates: differences in callus formation.
OBJECTIVES This study compared callus formation in distal femur fractures stabilized with locking plates and intramedullary nails to test the hypothesis that locking plates induce less fracture callus than IM nails. DESIGN Retrospective case matched study. SETTING Two orthopaedic trauma centers. PATIENTS 174 distal femur fracture were reviewed to extract cases treated with retrograde IM n...
متن کاملUse of locking plate in combination with dynamic compression plate for repair of tibial fracture in a young horse
Grade 1 open spiral fracture of left tibia in a 1.5 year-old horse was treated using a combination of 4.5 mm locking compression plate (LCP) and a dynamic compression plate applied in a neutralization manner. The open wound was managed postoperatively by regular dressing and lavage. Radiograph at 24 days post-treatment revealed callus formation and at two months complete bridging callus was obs...
متن کاملDynamic plate osteosynthesis for fracture stabilization: how to do it
Plate osteosynthesis is one treatment option for the stabilization of long bones. It is widely accepted to achieve bone healing with a dynamic and biological fixation where the perfusion of the bone is left intact and micromotion at the fracture gap is allowed. The indications for a dynamic plate osteosynthesis include distal tibial and femoral fractures, some midshaft fractures, and adolescent...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. American volume
دوره 98 6 شماره
صفحات -
تاریخ انتشار 2016