AIM The aim of our study was the dosimetric and physical evaluation of the CK and IMRT treatment plans for 16 patients with localized prostate cancer. BACKGROUND Intensity modulated radiation therapy (IMRT) is one of the recent technical advances in radiotherapy. The prostate is a well suited site to be treated with IMRT. The challenge of accurately delivering the IMRT needs to be supported by new advances such as image-guidance and four-dimensional computed conformal radiation therapy (4DCRT) tomography. CyberKnife (CK) provides real time orthogonal X-ray imaging of the patient during treatment course to follow gold fiducials installed into the prostate and to achieve motion correlation between online acquired X-ray imaging and digital reconstructed radiographs (DRRs) which are obtained from planning computed tomography images by translating and rotating the treatment table in five directions. METHODS AND MATERIALS Sixteen IMRT and CK plans were performed to be compared in terms of conformity (CI), heterogeneity indices (HI), percentage doses of 100% (V100), 66% (V66), 50% (V50), 33% (V33) and 10% (V10) volumes of the bladder and rectum. Dose-volume histograms for target and critical organs, (CI) and indices (HI) and isodose lines were analyzed to evaluate the treatment plans. RESULTS Statistically significant differences in the percentage rectal doses delivered to V10, V33, and V50 of the rectum were detected in favor of the CK plans (p values; <0.001, <0.001 and 0.019, respectively). The percentage doses for V66 and V100 of the rectum were larger in CK plans (13%, 2% in IMRT and 21%, 3% in CK plans, respectively). Percentage bladder doses for V10 and V33 were significantly lower in CK plans [96% in IMRT vs 48% in CK (p < 0.001) and 34% in IMRT vs 24% in CK (p = 0.047)]. Lower percentage doses were observed for V50, V66 of the bladder for the IMRT. They were 5.4% and 3.45% for IMRT and 13.4% and 8.05% for CK, respectively. Median CI of planning target volume (PTV) for IMRT and CK plans were 0.94 and 1.23, respectively (p < 0.001). CONCLUSION Both systems have a very good ability to create highly conformal volumetric dose distributions. Median HI of PTV for IMRT and CK plans were 1.08 and 1.33, respectively (p < 0.001).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

3-Dimensional conformal radiotherapy versus intensity modulated radiotherapy for localized prostate cancer: Dosimetric and radiobiologic analysis

 Background: To analyze the dosimetric and radio biologic advantages between intensity modulated radiotherapy (IMRT) and 3 dimensional conformal radiotherapy (3DCRT) and selection of optimal photon energy for IMRT treatments. Material and methods: 24 patients with localized prostate carcinoma were planned for 3DCRT and IMRT techniques. Radiation dose of 54 Gy with 2 Gy/fraction, was planned to ...

متن کامل

Radiation treatment planning for prostate cancer: A new dosimetric comparison of five and seven fields IMRT plans

Background: To compare the dosimetric coverage of the planning target volume (PTV) and the dose delivered to the main Organs at Risk (OARs) in 5 and 7-field techniques of Intensity Modulated Radiation Therapy (IMRT) in patients with local prostate cancer. Materials and Methods: Twelve patients with local prostate cancer underwent 5 and 7-field IMRT planning. The delivery of IMRT was carried out...

متن کامل

The Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions

Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans. Materials and Method...

متن کامل

Dosimetric Comparison of Various Radiotherapeutic Treatment Plans Using Cobalt-60 and Linac for Prostate Cancer

Introduction: This study was designed to evaluate dosimetric parameters such as rectal dose, bladder dose, integral dose (ID), homogeneity index (HI) and conformity index (CI) in various treatment plans for prostate cancer. Materials and Methods: Pelvic computed tomography (CT) scans of 27 patients were acquired and transferred to the RTDosePLAN treatment planning system. For each patient, 8 co...

متن کامل

Comparison of Radiobiological Models for Radiation Therapy Plans of Prostate Cancer: Three-dimensional Conformal versus Intensity Modulated Radiation Therapy

Purpose: In the current study, using different radiobiological models, tumor control probability (TCP) and normal tissue complication probability (NTCP) of radiotherapy plans were calculated for three-dimensional conformal radiation therapy (3D-CRT) and intensity modulated radiation therapy (IMRT) of prostate cancer.Methods and Materials: 10 prostate plans were randomly selected among patients ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology

دوره 15 6  شماره 

صفحات  -

تاریخ انتشار 2010