Reply on "What Is the Role of Surgeons When Implanting a Totally Implantable Venous Access Device to Prevent Immediate Complications?"
نویسندگان
چکیده
To the editor: We appreciate your great interest in our article [1]. First of all, I agree with the comments concerning the risk of a blind technique and the need to check the catheter’s position during the procedure. In terms of safety, a surgical cut-down procedure on the cephalic vein or the external jugular vein is excellent, especially for avoiding fatal complications such as a pneumoor a hemothorax [2-4]. Also, ultrasonography guidance makes the procedure safer [5, 6]. I think that the main issue is whether or not ultrasonography or fluoroscopy is available in the operation room. In fact, in the past, neither ultrasonography nor fluoroscopy was freely available in the operation rooms at Konkuk University Medical Center. However, recently with the availability of these medical devices, ultrasonography is routinely used to insert the totally implantable central venous access devices (TICVAD) and to check the position of the catheter tip during the procedure, as you recommended. As to the terminology of port migration, your comment is correct, and changing “port migration” to “dislocation/ dislodgement of port chamber” would be much clearer, as you mentioned [7]. Additionally, I would like to comment on the classification of complication according to the period. The standard of early and late complications is different in different reports in the literature. Knebel et al. [7] classified complications as perioperative and postoperative. You have classified them as immediate, early and late complications or as early and late complications [2, 4]. As you have indicated in your comments, complications within 24 hours after the procedure are important because they are related to the surgeon or to technical failures. We were interested in the surgeon’s influence on the complications. Therefore, we wanted to focus on complications within 24 hours after the surgery, which we defined as early complications. As to your comment on heparin usage, we indicated the usage of 10 mL of diluted heparin. This means that a total of 10 mL of diluted heparin solution was used to flush and fill the catheter before and after TICVAD insertion into the central vein. Also, we used a solution of heparin sodium (500 IU of heparin in 10 mL of isotonic saline) for flushing.
منابع مشابه
Commentary on "Insertion of Totally Implantable Central Venous Access Devices by Surgeons" - What Is the Role of Surgeons When Implanting a Totally Implantable Venous Access Device to Prevent Immediate Complications?
Adriana Toro, Gaetano Bertino, Annalisa Ardiri, Isidoro Di Carlo Department of Surgery, Patti Hospital (ME), Patti; Hepatology Unit, Department of Medical and Pediatric Sciences, University of Catania, Catania; Department of Surgical Sciences and Advanced Technologies “G.F. Ingrassia”, University of Catania, Catania, Italy; Department of General Surgery, Hamad General Hospital, Doha, Qatar Lett...
متن کاملApplying Totally Implantable Venous Access Devices (TIVAD) in Children: the First Iranian Experience
متن کامل
Implantable Port Devices, Complications and outcome in Pediatric Cancer, a Retrospective Study
Abstract Background Peripheral blood vessels, due to availability are used for many years in cancer patients, however in patients with potentially harmful drugs to skin (vesicant drugs) or difficult accessibility to vessels, the use of implantable port (totally implantable venous access port-TIVAP)devices with central vascular access are important. ...
متن کاملEffectiveness of Permanent Implantable Catheter (Polysite) in Children with Cancer
Abstract Background Totally implantable central venous access devices (ports) have been available for over 10 years, but have not been achieved widespread use in pediatric oncology patients. Ports facilitate the administration of chemotherapy in children with cancer. Materials and Methods In this study, early complications of implantable central venous access devices in children with diff...
متن کاملIntrapulmonal dislocation of a totally implantable venous access device
BACKGROUND: Totally implantable venous access devices are widely used for infusion of chemotherapy or parenteral nutrition. Device associated complications include technical operative problems, infections, paravasal infusions and catheter or punction chamber dislocation. CASE PRESENTATION: We present the case of a 49-year-old patient with the rare complication of a intrapulmonal catheter disloc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 31 شماره
صفحات -