Determination of the accuracy of implant reconstruction and dose delivery in brachytherapy in The Netherlands and Belgium.

نویسندگان

  • R J Elfrink
  • I K Kolkman-Deurloo
  • H J van Kleffens
  • A Rijnders
  • B Schaeken
  • T H Aalbers
  • W J Dries
  • J L Venselaar
چکیده

PURPOSE To gain insight into the accuracy of brachytherapy treatments, the accuracy of implant reconstruction and dose delivery was investigated in 33 radiotherapy institutions in The Netherlands and Belgium. MATERIALS AND METHODS The accuracy of the implant reconstruction method was determined using a cubic phantom containing 25 spheres at well-known positions. Reconstruction measurements were obtained on 41 brachytherapy localizers, 33 of which were simulators. The reconstructed distances between the spheres were compared with the true distances. The accuracy of the dose delivery was determined for high dose rate (HDR), pulsed dose rate (PDR) and low dose rate (LDR) afterloading systems using a polymethyl methacrylate cylindrical phantom containing a NE 2571 ionization chamber in its centre. The institutions were asked to deliver a prescribed dose at the centre of the phantom. The measured dose was compared with the prescribed dose. RESULTS The average reconstruction accuracy was -0.07 mm (+/-0.4 mm, 1 SD) for 41 localizers. The average deviation of the measured dose from the prescribed dose was +0.9% (+/-1.3%, 1 SD) for 21 HDR afterloading systems, +1.0% (+/-2.3%, 1 SD) for 12 PDR afterloaders, and +1.8% (+/-2.5%, 1 SD) for 15 LDR afterloaders. CONCLUSIONS This comparison showed a good accuracy of brachytherapy implant reconstruction and dose delivery in The Netherlands and Belgium.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A study on dosimetry accuracy of Strut-Adjust Volume Implant (SAVI) brachytherapy

Introduction: Accelerated Partial Breast Irradiation (APBI) is an effective treatment for breast carcinomas because of its limited number of fractions. Strut-adjusted volume implant (SAVI) is a HDR brachytherapy applicator to deliver the prescription dose with good PTV coverage and acceptable dose to skin, chest wall and organ at risks. Acceptable clinical outcomes depend on a...

متن کامل

Developing a Verification and Training Phantom for Gynecological Brachytherapy System

Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...

متن کامل

Modification of Source Strength in Low-Dose-Rate Lung Brachytherapy with 125I and 103Pd seeds

Background: A new treatment approach for most patients who have undergone early stage non-small-cell lung carcinoma (NSCLC), is wedge resection plus permanent implant brachytherapy. However, the specification of dose to medium at low energies especially in heterogeneous lung is unclear yet.Objective: The present study aims to modify source strength for different configurations of 125I and 103...

متن کامل

Comparison and Evaluation of the Effects of Rib and Lung Inhomogeneities on Lung Dose in Breast Brachytherapy using a Treatment Planning System and the MCNPX Code

Introduction: This study investigates to what extent the computed dose received by lung tissue in a commercially available treatment planning system (TPS) for 192Ir high-dose-rate breast brachytherapy is accurate in view of tissue inhomogeneities and presence of ribs. Materials and Methods: A CT scan of the breast was used to construct a patient-equivalent phantom in the clinical treatment plan...

متن کامل

Updating the Planar Patterson-Parker Table for Ir-192 and Cs-137 Brachytherapy Sources Using the Most Recent TG-43U1 Recommended Dosimetric Parameters

Background: The Patterson-Parker table was created in 1934 to determine mg-hr required to deliver the prescribed dose to the treatment area. These tables were created using the dosimetric data for 226Ra that was determined in air and were utilized for dose calculations around implants with Ra-equivalent radionuclides such as 137Cs and 192Ir. Therefore, the differences of the tissue attenuation ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology

دوره 59 3  شماره 

صفحات  -

تاریخ انتشار 2001