Accuracy of Four Blood Glucose Monitoring Systems
نویسندگان
چکیده
Method: Accuracy of the four blood glucose monitoring systems for fingertip capillary blood testing was assessed at three diabetes clinics. A total of 142 diabetic subjects were included in the study. At each study site, the trained operator tested the subject’s fingertip blood with the four systems and a YSI glucose analyzer, which served as the reference. Accuracy was evaluated according to ISO 15197 by calculating the percentage of meter results falling within ±5%, ±10%, ±15% and ±20% of the reference value for glucose concentrations 75 mg/dL (4.2 mmol/L) or higher, and within ±5, ±10, ±15 mg/dL (±0.28, ±0.56 and ±0.83 mmol/L) of the reference value for glucose concentrations below 75 mg/dL (4.2 mmol/L).
منابع مشابه
Accuracy of Four Blood Glucose Monitoring Systems
Background: The minimum acceptable accuracy for results produced by a blood glucose monitoring system according to the current version of ISO 15197, created in 2003, is: 95% of the individual glucose results shall fall within ±15 mg/dL (0.83 mmol/L) of the results of the manufacturer’s measurement procedure at glucose concentrations < 75 mg/dL (< 4.2 mmol/L) and within ±20% at glucose concentra...
متن کاملCan the Accuracy of Home Blood Glucose Monitors be affected by the Received Signal Strength of 900 MHz GSM Mobile Phones?
Background: People who use home blood glucose monitors may use their mobile phones in the close vicinity of medical devices. This study is aimed at investigating the effect of the signal strength of 900 MHz GSM mobile phones on the accuracy of home blood glucose monitors.Methods: Sixty non-diabetic volunteer individuals aged 21 - 28 years participated in this study. Blood samples were analyzed ...
متن کاملElectromagnetic Radiofrequency Radiation Emitted from GSM Mobile Phones Decreases the Accuracy of Home Blood Glucose Monitors
Mobile phones are two-way radios that emit electromagnetic radiation in microwave range. As the number of mobile phone users has reached 6 billion, the bioeffects of exposure to mobile phone radiation and mobile phone electromagnetic interference with electronic equipment have received more attention, globally. As self-monitoring of blood glucose can be a beneficial part of diabetes control, ho...
متن کاملAccuracy evaluation of five blood glucose monitoring systems: the North American comparator trial.
BACKGROUND This study evaluated differences in accuracy between the CONTOUR® NEXT EZ (EZ) blood glucose monitoring system (BGMS) and four other BGMSs [ACCU-CHEK® Aviva (ACAP), FreeStyle Freedom Lite® (FFL), ONE TOUCH® Ultra®2 (OTU2), and TRUEtrack® (TT)]. METHODS Up to three capillary blood samples (N = 393) were collected from 146 subjects with and without diabetes. One sample per subject wa...
متن کاملEvaluation of accuracy, precision and consensus of four laboratory glucose measurement kits with reference method
Abstract Background and Objective: According to recent changes in diagnostic criteria for diabetes, the harmonization of results obtained from various methods and systems by considering their accuracy and precision is essential. This study aimed to evaluate the accuracy, precision and consensus of some routine laboratory glucose kits in co...