Characterization of a hyper-viscoelastic phantom mimicking biological soft tissue using an abdominal pneumatic driver with magnetic resonance elastography (MRE).

نویسندگان

  • Gwladys E Leclerc
  • Laëtitia Debernard
  • Félix Foucart
  • Ludovic Robert
  • Kay M Pelletier
  • Fabrice Charleux
  • Richard Ehman
  • Marie-Christine Ho Ba Tho
  • Sabine F Bensamoun
چکیده

The purpose of this study was to create a polymer phantom mimicking the mechanical properties of soft tissues using experimental tests and rheological models. Multifrequency Magnetic Resonance Elastography (MMRE) tests were performed on the present phantom with a pneumatic driver to characterize the viscoelastic (μ, η) properties using Voigt, Maxwell, Zener and Springpot models. To optimize the MMRE protocol, the driver behavior was analyzed with a vibrometer. Moreover, the hyperelastic properties of the phantom were determined using compressive tests and Mooney-Rivlin model. The range of frequency to be used with the round driver was found between 60 Hz and 100 Hz as it exhibits one type of vibration mode for the membrane. MRE analysis revealed an increase in the shear modulus with frequency reflecting the viscoelastic properties of the phantom showing similar characteristic of soft tissues. Rheological results demonstrated that Springpot model better revealed the viscoelastic properties (μ=3.45 kPa, η=6.17 Pas) of the phantom and the Mooney-Rivlin coefficients were C(10)=1.09.10(-2) MPa and C(01)=-8.96.10(-3) MPa corresponding to μ=3.95 kPa. These studies suggest that the phantom, mimicking soft tissue, could be used for preliminary MRE tests to identify the optimal parameters necessary for in vivo investigations. Further developments of the phantom may allow clinicians to more accurately mimic healthy and pathological soft tissues using MRE.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Viscoelastic Modeling of Brain MRE Data Using FE Method

Dynamic shear test on simulated phantom can be used to validate magnetic resonance elastography (MRE) measurements. Phantom gel has been usually utilized for the cell culture of cartilage and soft tissue and also been used for mechanical property characterization using imaging systems. The viscoelastic property of the phantom would be important for dynamic experiments and analyses. In this stud...

متن کامل

Ergonomic Flexible Drivers for Hepatic MR Elastography

Introduction: Normal liver tissue is soft when assessed by palpation, whereas cirrhotic livers are typically very hard [1]. Liver stiffness is a very sensitive and specific biomarker for staging liver fibrosis in chronic liver disease patients, which has been proved by a number of elastographic techniques such as ultrasound elasticity imaging and Magnetic Resonance Elastography (MRE) [2-6]. Eme...

متن کامل

Focused Acoustic Driver to Generate High-Frequency Shear Waves in Deep Regions for Magnetic Resonance Elastography

INTRODUCTION Magnetic resonance elastography (MRE) is a phase-contrast technique that can noninvasively visualize shear waves patterns within tissue [1]. Accurate values of tissue stiffness are calculated by the equations of motion from the shear wave pattern of a single frequency [2]. In order to generate shear waves within tissue, various external drivers have been proposed [3]. To acquire an...

متن کامل

I-51: Pioneering New Approaches to Reproductive Imaging

Background: Gynecologic conditions including uterine fibroids, adenomyosis and endometriosis lead to significant impairment in women’s’ lives and significant expen- Abstracts of the 12th Royan International Congress on Reproductive Biomedicine International Journal of Fertility & Sterility (IJFS), Vol 5, Suppl 1, Summer 2011 24 ditures of health care dollars. However, many women have symptoms l...

متن کامل

Identification and quantification of myofascial taut bands with magnetic resonance elastography.

OBJECTIVE To explore the feasibility of using a new magnetic resonance imaging (MRI) technique--magnetic resonance elastography (MRE)--to identify and quantitate the nature of myofascial taut bands. DESIGN This investigation consisted of 3 steps. The first involved proof of concept on gel phantoms, the second involved numeric modeling, and the third involved a pilot trial on 2 subjects. Imagi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of biomechanics

دوره 45 6  شماره 

صفحات  -

تاریخ انتشار 2012